Shim-pack IC-C1/IC-GC1
Manuals | 2025 | ShimadzuInstrumentation
The analysis of inorganic cations is critical in environmental monitoring, water quality control, pharmaceutical testing and industrial process control. High‐performance cation exchange columns such as Shim-pack IC-C1 enable rapid, precise separation of alkali metals, alkaline earth metals, transition metal ions and alkylamines using non-suppressor ion chromatography. Their robustness, pH tolerance and compatibility with HPLC systems make them indispensable tools in modern analytical laboratories.
This document provides a comprehensive instruction manual for the Shim-pack IC-C1 analytical column and its guard column Shim-pack IC-GC1. It outlines specifications, installation procedures, eluent selection, operating conditions, sample preparation guidelines, cleaning protocols, storage recommendations and disposal considerations to ensure reproducible performance and long column lifetime.
The Shim-pack IC-C1 is packed with a styrene-divinylbenzene copolymer functionalized with sulfo groups (10 µm particle size) optimized for cation exchange in a non-suppressor conductivity detection mode. Key operational parameters include:
Eluent choices depend on ion valence:
Upon shipment, each Shim-pack IC-C1 meets the following performance specification for sodium ions: theoretical plates (N) > 2,500 at 1.5 mL/min, 40 °C, 5 mmol/L HNO₃, with 25 µL of 10 mg/L Na⁺, detected by conductivity. Retention times and peak shapes may vary with usage and sample composition. A guard column installed upstream extends analytical column life and traps impurities. Proper installation—observing flow direction, minimizing dead volume and using recommended tubing—ensures optimal resolution and low backpressure.
Cleaning protocols using 20 mmol/L ethylenediamine with 40 mmol/L tartaric acid at ~0.5 mL/min for 1–2 hours can recover performance when hydrophobic contaminants or multivalent ions cause pressure rise or retention shifts. Separate cleaning of guard and analytical columns is advised. After cleaning, flush thoroughly with the working eluent.
Storage recommendations include capping both ends with stop-plugs, storing at ambient temperature away from extremes, and replacing the eluent monthly if unused beyond one month. Avoid drying the packing material to prevent irreversible damage.
Shim-pack IC-C1 offers:
Future developments may include smaller particle sizes or monolithic stationary phases to increase throughput and resolution, integrated suppressor designs for enhanced sensitivity, hyphenation with mass spectrometry for speciation analysis, automated guard column exchange and greener eluents to reduce environmental impact. Advances in column materials may also enable separation of complex matrices such as emerging contaminants and biometals.
This instruction manual for Shim-pack IC-C1 and its guard column provides essential guidelines for installation, operation, maintenance and storage to achieve reproducible ion chromatographic separations. Adherence to recommended conditions ensures high performance and long service life in diverse analytical applications.
No literature references were provided in the original text.
Consumables, LC columns, Ion chromatography
IndustriesManufacturerShimadzu
Summary
Significance of the Topic
The analysis of inorganic cations is critical in environmental monitoring, water quality control, pharmaceutical testing and industrial process control. High‐performance cation exchange columns such as Shim-pack IC-C1 enable rapid, precise separation of alkali metals, alkaline earth metals, transition metal ions and alkylamines using non-suppressor ion chromatography. Their robustness, pH tolerance and compatibility with HPLC systems make them indispensable tools in modern analytical laboratories.
Objectives and Study Overview
This document provides a comprehensive instruction manual for the Shim-pack IC-C1 analytical column and its guard column Shim-pack IC-GC1. It outlines specifications, installation procedures, eluent selection, operating conditions, sample preparation guidelines, cleaning protocols, storage recommendations and disposal considerations to ensure reproducible performance and long column lifetime.
Methodology
The Shim-pack IC-C1 is packed with a styrene-divinylbenzene copolymer functionalized with sulfo groups (10 µm particle size) optimized for cation exchange in a non-suppressor conductivity detection mode. Key operational parameters include:
- Column dimensions: 150 mm × 5.0 mm I.D. (analytical) and 10 mm × 4.0 mm I.D. (guard)
- Recommended flow rate: 1.5 mL/min (maximum 2.0 mL/min)
- Operating temperature: 40 °C (maximum 50 °C)
- Pressure limit: below 2.0 MPa (maximum 5.0 MPa)
- pH range: 2–11
- Organic modifier: up to 10% methanol or acetonitrile
Eluent choices depend on ion valence:
- Monovalent cations: 5 mmol/L nitric acid aqueous solution
- Multivalent cations: 4 mmol/L tartaric acid with 2 mmol/L ethylenediamine (adjustable to 1 mmol/L if retention is weak)
Used Instrumentation
- HPLC system equipped with a non-suppressor conductivity detector
- Shim-pack IC-C1 analytical column and Shim-pack IC-GC1 guard column
- PEEK tubing (0.25–0.3 mm I.D., 1.6 mm O.D.) and PEEK male nuts
- Optional pre-heating coil kit (P/N 228-45714-91) between injector and column
Main Results and Discussion
Upon shipment, each Shim-pack IC-C1 meets the following performance specification for sodium ions: theoretical plates (N) > 2,500 at 1.5 mL/min, 40 °C, 5 mmol/L HNO₃, with 25 µL of 10 mg/L Na⁺, detected by conductivity. Retention times and peak shapes may vary with usage and sample composition. A guard column installed upstream extends analytical column life and traps impurities. Proper installation—observing flow direction, minimizing dead volume and using recommended tubing—ensures optimal resolution and low backpressure.
Cleaning protocols using 20 mmol/L ethylenediamine with 40 mmol/L tartaric acid at ~0.5 mL/min for 1–2 hours can recover performance when hydrophobic contaminants or multivalent ions cause pressure rise or retention shifts. Separate cleaning of guard and analytical columns is advised. After cleaning, flush thoroughly with the working eluent.
Storage recommendations include capping both ends with stop-plugs, storing at ambient temperature away from extremes, and replacing the eluent monthly if unused beyond one month. Avoid drying the packing material to prevent irreversible damage.
Benefits and Practical Applications
Shim-pack IC-C1 offers:
- High selectivity and efficiency for a broad range of cations
- Wide pH compatibility (2–11) and tolerance to low organic content
- Robust performance under routine QA/QC, environmental and industrial analyses
- Ease of maintenance via guard column replacement and simple cleaning
Future Trends and Potential Applications
Future developments may include smaller particle sizes or monolithic stationary phases to increase throughput and resolution, integrated suppressor designs for enhanced sensitivity, hyphenation with mass spectrometry for speciation analysis, automated guard column exchange and greener eluents to reduce environmental impact. Advances in column materials may also enable separation of complex matrices such as emerging contaminants and biometals.
Conclusion
This instruction manual for Shim-pack IC-C1 and its guard column provides essential guidelines for installation, operation, maintenance and storage to achieve reproducible ion chromatographic separations. Adherence to recommended conditions ensures high performance and long service life in diverse analytical applications.
References
No literature references were provided in the original text.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Shim-pack IC-A1/IC-GA1
2025|Shimadzu|Manuals
228-10327A Jan. 2025 High Performance Packed Column for HPLC Shim-pack IC-A1/IC-GA1 INSTRUCTION MANUAL 1. Introduction The Shim-pack IC-A1/GA1 is an ion chromatography column filled with a polymer gel that contains anionic exchange groups. The Shimpack IC-A1 is an analytical column…
Key words
カラム, カラムイオン, イオンガードカラム, ガードカラムメイルナット, メイルナットイオンクロマトグラフィー, イオンクロマトグラフィーピーク, ピークメタノール, メタノールcleaning, cleaningpeek, peekアミノメタン, アミノメタンメンブランフィルター, メンブランフィルターノンサプレッサ, ノンサプレッササイズ, サイズグレード, グレードcolumn
Shimadzu IC-SA4 HPLC Column
|Shimadzu|Manuals
228-91805A Jul.2023 Shimadzu High Performance Packed Column for HPLC Shim-pack IC-SA4 Instruction Manual NOTE: Compound retention times and peak shape may vary with usage. Before developing an analytical method with the column, verify that ■ Introduction Shim-pack IC-SA4 is a…
Key words
カラム, カラムイオン, イオンサプレッサ, サプレッサピーク, ピークガードフィルタ, ガードフィルタイオンクロマトグラフィー, イオンクロマトグラフィーメイルナット, メイルナットcolumn, columnpeek, peekflushing, flushingguard, guardexit, exitwaste, wastemay, mayパッケー
Shimadzu IC-A3/-GA3 HPLC Column
|Shimadzu|Manuals
228-30003D Feb. 2009 High Performance Packed Column for High Performance Liquid Chromatograph Shim-pack IC-A3/-GA3 Instruction Manual Introduction Shim-pack IC-A3 is an polymer gel based column with an anion exchange group for ion chromatography. It is optimized for non-suppressed ion…
Key words
カラム, カラムイオン, イオンガードカラム, ガードカラムピーク, ピークcolumn, columnイオンクロマトグラフィー, イオンクロマトグラフィーpeek, peekメイルナット, メイルナットエタノール, エタノールメタノール, メタノールノンサプレッサ, ノンサプレッサシリーズ, シリーズグレード, グレードアセトニトリル, アセトニトリルflushing
Shim-pack NovaCore C18-HB [Reversed phase column]
|Shimadzu|Manuals
日本語取扱説明書はP.4以降をご参照ください。 ERAD-0001-9087A High Performance Packed Column for HPLC Shim-pack NovaCore C18-HB [Reversed phase column] 取扱説明 書 INSTRUCTION MANUAL ■ Introduction NOTE Near the pH limit of use, please use a mobile phase containing at least 10% organic solvent. At the…
Key words
カラム, カラムイオンペア, イオンペアシステム, システムメイルナット, メイルナットピーク, ピークcolumn, columnthf, thfsus, susノイズ, ノイズベースラインドリフト, ベースラインドリフトprecautions, precautionsmobile, mobileチューブ, チューブphase, phaseflow