Simplified approach for structural elucidation and quantitation for pharmaceutical API and related impurities
Posters | 2019 | ShimadzuInstrumentation
Structural characterization and quantification of impurities in pharmaceutical active ingredients are critical to ensure drug safety, efficacy, and regulatory compliance. Rapid and reliable workflows that combine high-resolution mass spectrometry with advanced data analysis tools help laboratories identify unknown degradation products and maintain product quality.
This study introduces a streamlined strategy for elucidating the structures of paracetamol impurities and performing quantitative analysis on both the parent drug and related compounds using Shimadzu’s LCMS-9030 Q-TOF system paired with Insight Explore software.
Sample Preparation:
LC–MS/MS Conditions:
Mass Spectrometer: LCMS-9030 (Q-TOF) with heated ESI and CDS probe; features include UFgrating, iRefTOF tube, 100 Hz acquisition, mass range m/z 10–40 000, resolution >30 000, and mass accuracy <1 ppm.
Data Analysis:
Known Impurities:
Unknown Impurities:
Quantitation:
By integrating high-resolution Q-TOF performance with Insight Explore software, the workflow reduces reliance on multiple analytical techniques while delivering confident structural assignments. Laboratories can achieve both identification and quantitation in a single run, improving throughput in R&D and quality control environments.
Advances in instrument speed, resolution, and data processing will further streamline impurity profiling. Integration with machine learning for automated structural proposals, expanded compound libraries, and standardized reporting formats will accelerate method development and regulatory submissions.
The combined use of LCMS-9030 Q-TOF and Insight Explore offers a robust platform for rapid structural elucidation and accurate quantitation of paracetamol and its related impurities. The approach ensures mass accuracy below 2 ppm, reliable formula assignment, and excellent calibration linearity, supporting comprehensive impurity analysis in pharmaceutical workflows.
LC/TOF, LC/HRMS, LC/MS, LC/MS/MS
IndustriesPharma & Biopharma
ManufacturerShimadzu
Summary
Importance of the Topic
Structural characterization and quantification of impurities in pharmaceutical active ingredients are critical to ensure drug safety, efficacy, and regulatory compliance. Rapid and reliable workflows that combine high-resolution mass spectrometry with advanced data analysis tools help laboratories identify unknown degradation products and maintain product quality.
Objectives and Overview
This study introduces a streamlined strategy for elucidating the structures of paracetamol impurities and performing quantitative analysis on both the parent drug and related compounds using Shimadzu’s LCMS-9030 Q-TOF system paired with Insight Explore software.
Methodology and Instrumentation
Sample Preparation:
- Paracetamol reference and sample solutions: 50 mg in 5 mL water/methanol (85:15 v/v).
- Impurity stock solutions: 5 mg impurities in 50 mL diluent; further diluted for calibration levels.
LC–MS/MS Conditions:
- UHPLC: Nexera X2 with Shim-pack XR-ODSII column (1.5 mm×100 mm) at 30 °C.
- Mobile phase: 20 mM ammonium acetate (A) and acetonitrile (B); gradient over 30 min at 0.3 mL/min.
- Detection wavelength: 361 nm, injection volume: 1 µL.
Mass Spectrometer: LCMS-9030 (Q-TOF) with heated ESI and CDS probe; features include UFgrating, iRefTOF tube, 100 Hz acquisition, mass range m/z 10–40 000, resolution >30 000, and mass accuracy <1 ppm.
Data Analysis:
- Formula prediction via Insight Explore based on accurate precursor masses.
- Generation of fragmentation pathways and .mol files for structural proposals.
- Confirmation through matching experimental MS/MS fragments with theoretical masses.
Main Results and Discussion
Known Impurities:
- All standard paracetamol impurities (2, 4, J, K, L) yielded mass accuracies below 2 ppm.
- Insight Explore assigned correct molecular formulas with scores above 97%.
- MS/MS fragmentation patterns supported proposed structures.
Unknown Impurities:
- Three sample impurities detected at retention times ~4.3, 8.7, and 11.7 min.
- Accurate masses led to formula assignments (C12H12N2O, C16H16N2O3, C9H11NO2) with mass errors <1 ppm.
- Fragment analysis confirmed likely structural features for each unknown.
Quantitation:
- Calibration for paracetamol: 50–500 ppb (R²>0.998).
- Impurity J and impurity 4: 100–1000 ppb (R²>0.999).
- Demonstrated linear response and sensitivity for simultaneous qualitative and quantitative workflows.
Benefits and Practical Applications
By integrating high-resolution Q-TOF performance with Insight Explore software, the workflow reduces reliance on multiple analytical techniques while delivering confident structural assignments. Laboratories can achieve both identification and quantitation in a single run, improving throughput in R&D and quality control environments.
Future Trends and Possibilities
Advances in instrument speed, resolution, and data processing will further streamline impurity profiling. Integration with machine learning for automated structural proposals, expanded compound libraries, and standardized reporting formats will accelerate method development and regulatory submissions.
Conclusion
The combined use of LCMS-9030 Q-TOF and Insight Explore offers a robust platform for rapid structural elucidation and accurate quantitation of paracetamol and its related impurities. The approach ensures mass accuracy below 2 ppm, reliable formula assignment, and excellent calibration linearity, supporting comprehensive impurity analysis in pharmaceutical workflows.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Identification and Structural Elucidation of Amlodipine Impurities Using High Resolution LC/MS and LC-MS/MS
2015|Agilent Technologies|Applications
Identification and Structural Elucidation of Amlodipine Impurities Using High Resolution LC/MS and LC-MS/MS Application Note Pharmaceutical Authors Abstract Dinesh Udawant Characterization of impurities in drug products is an essential requirement in Anant Pharmaceuticals, the pharmaceutical development process. The identification of…
Key words
fragmentation, fragmentationmass, massamlodipine, amlodipinefragment, fragmentformula, formulacounts, countsmsc, msccharge, chargeimpurity, impuritysoftware, softwareelucidation, elucidationpcdl, pcdlstructure, structureaccurate, accuratestructures
Evaluation of the Importance of Accurate Mass, Mass Resolution and Dynamic Range for Impurity Profiling Applications with Multistage Mass Spectrometry
2008|Agilent Technologies|Presentations
Evaluation of the Importance of Accurate Mass, Mass Resolution and Dynamic Range for Impurity Profiling Applications with Multistage Mass Spectrometry David A. Weil*, Zoltan Timar and Michael Zumwalt Agilent Technologies Schaumburg, IL/ Boulder, CO [email protected] Impurity Profiling 2008 Page 1…
Key words
tof, tofmass, masscooh, coohqqq, qqqamoxicillin, amoxicillinmetid, metiddegradation, degradationtic, ticimpurity, impurityquad, quadfpga, fpgamfe, mfeprocessed, processeddata, datadynamic
Analysis of pharmaceuticals and drug related impurities using Agilent instrumentation
2010|Agilent Technologies|Guides
Analysis of pharmaceuticals and drug related impurities using Agilent instrumentation Application Compendium Introduction The analysis of impurities in drugsubstances, from initial screening to the use of validated methods in routine quality control and quality assurance, is becoming an increasingly challenging…
Key words
mass, massagilent, agilentformula, formulatime, timemau, maumin, minwere, wereamoxicillin, amoxicillinxylene, xylenemda, mdappm, ppmmethod, methodimpurities, impuritiesfrom, fromanalysis
Search for What’s Missing: Unknown Compound Characterization Using LC-MS
2019|Thermo Fisher Scientific|Presentations
Search for What’s Missing: Unknown Compound Characterization Using LC-MS Kate Comstock Sr. Marketing Specialist, Pharma/Biopharma Marketing The world leader in serving science Small Molecule Structure Analysis Small molecule structure analysis encompasses broad applications: Pharmaceutical, metabolomics, food & environmental, clinical, forensic,…
Key words
acquirex, acquirexmsn, msnhydrolysis, hydrolysisstructure, structureworkflow, workfloworbitrap, orbitrapamprenavir, amprenavirmass, massoxidation, oxidationmetabolites, metabolitesisotope, isotopedata, datamolecule, moleculeparent, parentmzcloud