Thermo Scientific Orbitrap Technology – Principle of Operation

Others |  | Thermo Fisher ScientificInstrumentation
LC/HRMS, LC/MS, LC/MS/MS, LC/Orbitrap
Industries
Manufacturer
Thermo Fisher Scientific

Summary

Importance of Orbitrap Technology


Orbitrap analyzers combine high resolution and mass accuracy with robust performance, making them vital for proteomics, metabolomics, and complex mixture analysis in both research and industrial settings.

Study Objectives and Overview


This document outlines the operational principles of Thermo Scientific Orbitrap technology, detailing ion handling from injection to detection. It aims to clarify the sequence of ion trapping, ejection, and signal processing steps that underpin the Orbitrap’s performance.

Methodology and Instrumentation


Major components:
  • Linear Ion Trap – initial trapping, isolation, and fragmentation of ions.
  • C-Trap – intermediate ion storage and collisional cooling using nitrogen gas to improve ion beam focus.
  • Orbitrap Analyzer – off-axis injection of ions, axial oscillation around a central electrode.
  • Detectors and Electronics – measurement of image currents on split outer electrodes; signal amplification and Fourier transformation for frequency-to-mass conversion.

Operational workflow:
  1. Ions are generated in the source and accumulated in the linear ion trap.
  2. Selected ions are ejected axially into the C-Trap and cooled via nitrogen collisions.
  3. Ions are transferred into the Orbitrap where increasing central electrode voltage induces tight orbital motion.
  4. Axial oscillation frequencies are recorded as image currents and converted to a mass spectrum by fast Fourier transformation.

Key Findings and Discussion


The Orbitrap’s mass accuracy derives from a direct relationship between ion oscillation frequency (ωz) and m/z ratio. Stable axial motion yields high resolving power exceeding 100,000 at m/z 200. Collisional cooling in the C-Trap enhances ion packet coherence, improving signal-to-noise ratio and dynamic range.

Benefits and Practical Applications


High mass accuracy and resolution enable:
  • Confident identification of peptides and small molecules.
  • Quantitative analysis in complex biological matrices.
  • Characterization of post-translational modifications.
  • Quality control in pharmaceutical and environmental laboratories.

Future Trends and Potential Applications


Advancements may include integration with ion mobility separation, real-time data acquisition algorithms, and AI-driven spectral interpretation. Miniaturization and ruggedization could extend Orbitrap use to field-deployable instruments.

Conclusion


Thermo Scientific Orbitrap technology delivers unmatched mass accuracy and resolution through precise ion manipulation and Fourier-based detection. Its versatile design supports a wide range of analytical challenges, promising continued impact across life sciences and industrial analytics.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
IMSC: Orbitrap Mass Spectrometry: from Dream to Mainstream
IMSC: Orbitrap Mass Spectrometry: from Dream to Mainstream
2009|Thermo Fisher Scientific|Presentations
Lunch talk at 18th IMSC Bremen Orbitrap Mass Spectrometry: from Dream to Mainstream Orbitrap is… …awarded… Dr. Alexander Makarov Thermo Fisher Scientific …produced… …commercialized… September 2, 2009 …developed… Irkutsk …in Bremen! Long time ago… A.Pekaln “If you really want to…
Key words
orbitrap, orbitrapions, ionsoatof, oatoftrap, trapion, ionoscillations, oscillationsmass, masstof, toffrequency, frequencyvoltage, voltageltq, ltqfield, fieldresolving, resolvingpartof, partoficr
High Capacity Electrostatic Ion Trap Mass Spectrometer and its Signal Processing
High Capacity Electrostatic Ion Trap Mass Spectrometer and its Signal Processing Design and Simulation On increasing charge capacity of EITs There have been increasing attempts to make use of larger space for ion trapping and measurement of the image charge…
Key words
field, fieldcharge, chargeimage, imagefrequency, frequencyeit, eitfft, fftion, ionfocusing, focusingisochronous, isochronousmotion, motionpitch, pitchaxsim, axsimtransform, transformrotational, rotationalprecession
Enhanced FT for Orbitrap Mass Spectrometry
Enhanced FT for Orbitrap Mass Spectrometry
2011|Thermo Fisher Scientific|Posters
Enhanced FT for Orbitrap Mass Spectrometry Oliver Lange, Eugen Damoc, Andreas Wieghaus, Alexander Makarov Thermo Fisher Scientific, Bremen, Germany Leff Δϕ = ω ⋅ tinj = The Fourier transformation of the time-domain transient provides a complex value for each point…
Key words
spectrum, spectrumabsorption, absorptionmagnitude, magnitudeeft, eftorbitrap, orbitrapimaginary, imaginaryspectra, spectraenhanced, enhancedapodized, apodizedunapodized, unapodizeddomain, domainmass, massmax, maxsin, sinpoint
Orbitrap Mass Spectrometry: Ultrahigh Resolution for Every Lab
Orbitrap Mass Spectrometry: Ultrahigh Resolution for Every Lab
2012|Thermo Fisher Scientific|Presentations
Symposium: New Alternatives in HighResolution Mass Spectrometry Orbitrap Mass Spectrometry: Ultrahigh Resolution for Every Lab Alexander Makarov March 14, 2012 What is Orbitrap™ analyzer? Orbitrap analyzer = = Orbital trapping + Image current detection + Electrodynamic squeezing + External pulsed…
Key words
orbitrap, orbitrapabundance, abundancerelative, relativehcd, hcdions, ionsanalyzer, analyzertrap, trapltq, ltqvelos, velospreamplifier, preamplifiermass, masshigh, hightic, ticmakarov, makarovfrequency
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike