High-throughput tandem capillary-flow LC-MS for maximum MS utilization
Technical notes | 2018 | Thermo Fisher ScientificInstrumentation
High-throughput proteomics demands methods that maximize mass spectrometer utilization while preserving sensitivity and coverage. Conventional nano-flow LC-MS offers excellent depth but suffers from idle MS time in fast turnover workflows. Capillary-flow tandem LC systems can bridge this gap, enabling continuous sample introduction and nearly 100 percent instrument usage in routine proteomic analyses.
This work aimed to develop and validate a tandem capillary-flow LC-MS approach that sustains uninterrupted peptide data acquisition. By configuring two parallel low-flow columns with staggered gradients and intelligent valve switching on a Thermo Scientific UltiMate 3000 RSLCnano platform, the study sought to reduce cycle times to seven minutes while achieving over 97 percent MS utility and robust proteome coverage.
Two identical Acclaim PepMap 100 C18 columns were installed in series with a second trap and post-column switching valve. While one column performs gradient elution, the second undergoes washing and equilibration. Look-ahead injections load the next sample onto the alternate trap during active elution. Key hardware and settings included
The tandem LC approach delivered continuous base peak chromatograms on each column with minimal dead time. Peptide elution spanned essentially the entire seven-minute runtime, translating into more than 200 samples per day at over 97 percent MS acquisition. Compared to a previous single-column eight-minute method, tandem operation increased MS/MS events and protein group identifications by 10 to 20 percent, despite broader peptide peak widths due to post-column dispersion.
Advancements may include integration of even faster gradients, automation of valve programs, and scaling to multiplexed assays. Tandem LC-MS could be adapted for quantitative label-free workflows in clinical and biopharma settings, as well as real-time monitoring in bioprocess control.
The described tandem capillary-flow LC-MS configuration achieves near-continuous MS utilization without sacrificing sensitivity or depth. It offers a practical route to unprecedented throughput in proteomics and can be tailored to diverse analytical demands.
1. Zhang X et al. A highly sensitive 150 µm column for high-throughput proteomics. Thermo Scientific App Note, 2018.
2. Boychenko A et al. High-throughput capillary-flow LC-MS proteomics with maximum MS utilization. Thermo Scientific Tech Note 72777, 2018.
3. Thermo Scientific UltiMate 3000 RSLCnano System Guide, 2016.
4. UltiMate 3000 RSLCnano Standard Applications Guide v3.0, 2018.
5. Complete guide to configuring Thermo Scientific nano LC for MS analysis.
6. Poster Note 64618. Human proteome profiling by label-free LC-MS, HUPO 2015.
LC/HRMS, LC/MS, LC/MS/MS, LC/Orbitrap
IndustriesManufacturerThermo Fisher Scientific
Summary
Importance of the Topic
High-throughput proteomics demands methods that maximize mass spectrometer utilization while preserving sensitivity and coverage. Conventional nano-flow LC-MS offers excellent depth but suffers from idle MS time in fast turnover workflows. Capillary-flow tandem LC systems can bridge this gap, enabling continuous sample introduction and nearly 100 percent instrument usage in routine proteomic analyses.
Objectives and Study Overview
This work aimed to develop and validate a tandem capillary-flow LC-MS approach that sustains uninterrupted peptide data acquisition. By configuring two parallel low-flow columns with staggered gradients and intelligent valve switching on a Thermo Scientific UltiMate 3000 RSLCnano platform, the study sought to reduce cycle times to seven minutes while achieving over 97 percent MS utility and robust proteome coverage.
Methodology and Instrumentation
Two identical Acclaim PepMap 100 C18 columns were installed in series with a second trap and post-column switching valve. While one column performs gradient elution, the second undergoes washing and equilibration. Look-ahead injections load the next sample onto the alternate trap during active elution. Key hardware and settings included
- UltiMate 3000 RSLCnano with NCS-3500RS and NCP-3200RS pumps
- Thermo Scientific Nanospray Flex source coupled to a Q Exactive HF-X Orbitrap
- Flow rates of 1.5 µL minute for gradients and 150 µL minute for loading
- Seven-minute cycles yielding a ~6.8-minute peptide window
Results and Discussion
The tandem LC approach delivered continuous base peak chromatograms on each column with minimal dead time. Peptide elution spanned essentially the entire seven-minute runtime, translating into more than 200 samples per day at over 97 percent MS acquisition. Compared to a previous single-column eight-minute method, tandem operation increased MS/MS events and protein group identifications by 10 to 20 percent, despite broader peptide peak widths due to post-column dispersion.
Benefits and Practical Applications
- Maximized instrument throughput for large sample cohorts
- Stable retention time reproducibility across rapid gradients
- Ease of implementation on existing RSLCnano platforms with minor fluidic modifications
- Applicability to routine QA/QC and discovery proteomics
Future Trends and Applications
Advancements may include integration of even faster gradients, automation of valve programs, and scaling to multiplexed assays. Tandem LC-MS could be adapted for quantitative label-free workflows in clinical and biopharma settings, as well as real-time monitoring in bioprocess control.
Conclusion
The described tandem capillary-flow LC-MS configuration achieves near-continuous MS utilization without sacrificing sensitivity or depth. It offers a practical route to unprecedented throughput in proteomics and can be tailored to diverse analytical demands.
References
1. Zhang X et al. A highly sensitive 150 µm column for high-throughput proteomics. Thermo Scientific App Note, 2018.
2. Boychenko A et al. High-throughput capillary-flow LC-MS proteomics with maximum MS utilization. Thermo Scientific Tech Note 72777, 2018.
3. Thermo Scientific UltiMate 3000 RSLCnano System Guide, 2016.
4. UltiMate 3000 RSLCnano Standard Applications Guide v3.0, 2018.
5. Complete guide to configuring Thermo Scientific nano LC for MS analysis.
6. Poster Note 64618. Human proteome profiling by label-free LC-MS, HUPO 2015.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Tandem nanoLC-MS: maximum MS utilization for deep-dive proteomic analysis
2019|Thermo Fisher Scientific|Technical notes
TECHNICAL NOTE 72899 Tandem nanoLC-MS: maximum MS utilization for deep-dive proteomic analysis Table of Contents Authors Christopher Pynn , Alexander Boychenko1, Wim Decrop1, Mike Baynham2, Peter Jehle1, Martin Ruehl1 1 Thermo Fisher Scientific, Germering, Germany 1 Thermo Fisher Scientific, Runcorn,…
Key words
loading, loadingpump, pumpnanolc, nanolctandem, tandemloop, loopnano, nanowash, washcomment, commentcolumn, columnvici, vicisample, samplepeptide, peptidevalve, valveproteomics, proteomicsflow
High-throughput capillary-flow LC-MS proteomics with maximum MS utilization
2018|Thermo Fisher Scientific|Technical notes
TECHNICAL NOTE 72777 High-throughput capillary-flow LC-MS proteomics with maximum MS utilization Authors Contents Alexander Boychenko1, Christopher Pynn1, Bart van den Berg1, Tabiwang N. Arrey2, Mike Baynham 3, Wim Decrop1, Martin Ruehl1 1. Introduction.............................................................................................. 2 Thermo Fisher Scientific, Germering, Germany 2.…
Key words
proteomics, proteomicsflow, flowthroughput, throughputtrap, trapthermo, thermodigest, digestscientific, scientificcolumn, columnhela, helawash, washhigh, highcapillary, capillaryequilibration, equilibrationproteome, proteomeloading
Maximize productivity in proteome profiling
2020|Thermo Fisher Scientific|Technical notes
TECHNICAL NOTE 73671 Maximize productivity in proteome profiling Combining intelligent tandem nanoLC-MS methods and a novel double-barrel ESI source for 100% MS utilization Authors: Runsheng Zheng1, Christopher Pynn1, Thomas Lanzinner2, Georg Völkle3, Jan Linnemann2, John Modrow2, Andreas Tebbe2, Wim Decrop1,…
Key words
nanolc, nanolcnanoviper, nanoviperscientific, scientificthermo, thermowashing, washingtandem, tandeminter, interloading, loadingpump, pumpemitter, emitterpulled, pulledcolumn, columnpacked, packedcolumns, columnssetup
Double Barrel ESI Source and Novel Tandem NanoLC-MS Setup Enables 24/7 Proteome Profiling with Close to 100% MS Utilization
2020|Thermo Fisher Scientific|Posters
Double Barrel ESI Source and Novel Tandem NanoLC-MS Setup Enables 24/7 Proteome Profiling with Close to 100% MS Utilization Runsheng Zheng1, Thomas Lanzinner2, Georg Völkle3, Christopher Pynn1, Jan Linnemann2, John Modrow2, Wim Decrop1, Andreas Tebbe2, Peter Jehle1, Mauro De Pra1,…
Key words
nanolc, nanolcwashing, washingtandem, tandemgradient, gradientloading, loadingplatform, platformnano, nanodeep, deephram, hramequilibration, equilibrationprofiling, profilingcolumns, columnstwosteps, twostepsproteome, proteomecolumn