LCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

Surviving Chromatography - Part 2: Corrective Action

Presentations | 2018 | Agilent TechnologiesInstrumentation
HPLC
Industries
Manufacturer
Agilent Technologies

Summary

Importance of the topic


Accurate and reproducible chromatography is vital for quality control, pharmaceutical development, environmental monitoring and research. Rapid identification and correction of HPLC performance issues preserve data integrity, reduce downtime and lower operational costs.

Objectives and Article Overview


This article reviews common HPLC problems—baseline instabilities, pressure fluctuations, peak shape anomalies and retention shifts—and proposes a systematic framework for diagnosing root causes and implementing corrective and preventive actions.

Methodology and Instrumentation


  • Systematic troubleshooting workflow: isolate column, mobile phase, sample and instrument as potential error sources.
  • Preventive maintenance practices: routine system flushing, inline filter replacement, sample filtration and proper mobile phase handling.
  • Key equipment: Agilent HPLC/UHPLC systems; Eclipse XDB-C8 and Zorbax RRHD columns; diode array (DAD) and refractive index (RID) detectors; autosamplers; inline filters; syringe filters.

Main Results and Discussion


  • Baseline Issues: Noise may originate from dirty flow cells, failing lamps or pump pulsation; drift often results from gradient effects, temperature instability or mobile phase contamination. Corrective measures include detector cleaning, lamp replacement, wavelength adjustment and mobile phase re-equilibration.
  • Pressure Anomalies: Sudden high backpressure typically indicates clogged inlet frits, guard columns or inline filters; low pressure suggests leaks or pump malfunctions. A stepwise pressure test—with and without column or guard cartridge—guides actions such as backflushing, filter replacement or column exchange.
  • Peak Shape Distortions: Tailing, splitting or broadening can arise from pH mismatch, inappropriate sample solvent, extra-column volume or hardware misalignment. Solutions involve adjusting buffer pH, matching injection solvent to mobile phase, minimizing dead volume and ensuring proper fitting connections.
  • Retention Shifts and Selectivity Changes: Small variations in mobile phase composition, pH, temperature or flow rate can alter retention times and selectivity. Consistent preparation protocols, accurate volumetric techniques and monitoring of critical parameters are essential to maintain method robustness.

Benefits and Practical Application


  • Enhances method robustness and reproducibility
  • Minimizes instrument downtime and maintenance expenses
  • Extends column lifespan and reduces sample carryover
  • Supports compliance with regulatory and quality assurance standards

Future Trends and Possibilities


Integration of smart diagnostics and AI-driven monitoring will enable real-time detection of performance drift. Emerging micro-flow and UHPLC technologies, advanced column materials and predictive maintenance algorithms will further streamline troubleshooting and enhance chromatographic efficiency.

Conclusion


A structured troubleshooting approach that evaluates column health, mobile phase quality, sample preparation and instrument condition ensures optimal HPLC performance. Routine preventive maintenance and adherence to best practices are key to sustained analytical success.

Reference


No external literature cited.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
The Chromatography Detective: Troubleshooting Tips  & Tools for LC & LCMS
The Chromatography Detective: Troubleshooting Tips & Tools for LC & LCMS Rita Steed LC Columns Application Engineer January 31, 2018 What do you do when..... ➢ Your chromatography changes • Peak shape • Retention ➢ You can’t reproduce a method…
Key words
beginner, beginnerpitfalls, pitfallsavoiding, avoidingcolumn, columndwell, dwellvolume, volumemobile, mobilephase, phasepeak, peakecv, ecvinstrument, instrumentshape, shapetailing, tailingretention, retentionconnections
Taking the Trouble Out of TroubleshootingIs It the Column, Method, or Instrument
Taking the Trouble Out of Troubleshooting Is It the Column, Method, or Instrument Rita Steed September 29, 2016 Page 1 What Do We Troubleshoot Typical LC troubleshooting approach asks: – What’s wrong with my column? – What’s wrong with my…
Key words
column, columnmobile, mobilephase, phaseproblems, problemsfrit, fritpeaks, peakstailing, tailingmau, mauplugged, pluggedfittings, fittingspressure, pressurepeak, peakmin, minshape, shapevolume
HPLC Column Troubleshooting: Is It Really The Column?
HPLC Column Troubleshooting: Is It Really The Column?
2010|Agilent Technologies|Presentations
HPLC Column Troubleshooting: Is It Really The Column? Agilent Technologies Technologies, Inc. Inc Rita Steed Application Engineer J January 22 22, 2010 Page 1 Agilent Restricted mA U Troubleshooting in HPLC 20 00 15 00 10 5 00 0 0…
Key words
restricted, restrictedagilent, agilentcolumn, columnmin, minmobile, mobiletailing, tailingtime, timephase, phasechange, changeretention, retentionplugged, pluggedplates, plateslot, lotchanges, changespeak
LC Column Troubleshooting - Isolating the Source of the Problem
LC Column Troubleshooting Isolating the Source of the Problem Rita Steed December 12, 2013 Page 1 What Do We Troubleshoot The typical LC troubleshooting approach asks the questions: – What’s wrong with the column? – What’s wrong with the instrument?…
Key words
column, columntailing, tailingmin, minmau, mauvolume, volumemobile, mobilefittings, fittingspeak, peakphase, phaseplugged, pluggedpeaks, peakstime, timefrit, fritred, redextra
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike