Exploring Factors That Influence Charge Variant Analysis of mAbs with Shim-pack Bio IEX Columns using Salt Gradient Elution
Applications | 2020 | ShimadzuInstrumentation
Monoclonal antibodies (mAbs) are a cornerstone of modern biopharmaceuticals, but their therapeutic efficacy and safety depend on precise characterization of charge variants arising from post-translational modifications. Ion-exchange chromatography (IEX) is the standard technique for resolving these variants, enabling quality control and consistency of mAb products.
This work evaluates the performance of Shim-pack Bio IEX SP-NP strong cation-exchange columns on a Nexera Bio UHPLC system for salt-gradient charge variant profiling of trastuzumab. Key goals include optimizing gradient slope, gradient time, flow rate, particle size, and column length to achieve high resolution and reproducible quantification of acidic, main and basic variants.
Sample Preparation and Chromatographic Conditions:
Instrumentation Used:
This optimized salt-gradient IEX method delivers a robust, high-resolution profiling of trastuzumab charge variants. It supports stringent QA/QC in biopharma development by providing clear separation of acidic and basic species, enabling accurate quantitation and monitoring of critical quality attributes.
Advances may include integration with mass spectrometric detection for structural characterization of variants, further miniaturization of columns to reduce solvent use, and automation of method scouting using machine learning to expedite method development across diverse mAbs.
The Shim-pack Bio IEX SP-NP columns coupled with a Nexera Bio UHPLC system present a flexible platform for salt-gradient charge variant analysis. Careful tuning of gradient slope, time, and column dimensions yields high-resolution separations, while flow rate and particle size can be adjusted to balance throughput and pressure constraints.
Consumables, HPLC, LC columns
IndustriesPharma & Biopharma
ManufacturerShimadzu
Summary
Significance of the Topic
Monoclonal antibodies (mAbs) are a cornerstone of modern biopharmaceuticals, but their therapeutic efficacy and safety depend on precise characterization of charge variants arising from post-translational modifications. Ion-exchange chromatography (IEX) is the standard technique for resolving these variants, enabling quality control and consistency of mAb products.
Study Objectives and Overview
This work evaluates the performance of Shim-pack Bio IEX SP-NP strong cation-exchange columns on a Nexera Bio UHPLC system for salt-gradient charge variant profiling of trastuzumab. Key goals include optimizing gradient slope, gradient time, flow rate, particle size, and column length to achieve high resolution and reproducible quantification of acidic, main and basic variants.
Methodology and Instrumentation
Sample Preparation and Chromatographic Conditions:
- Trastuzumab at 5 mg/mL in water, 2 µL injection.
- Mobile phases: A – 20 mM sodium phosphate buffer (pH 6.0); B – same buffer with 250 mM NaCl.
- Gradient elution from 0 to 250 mM NaCl.
Instrumentation Used:
- Shimadzu Nexera Bio UHPLC with non-metallic flow path.
- Shim-pack Bio IEX SP-NP columns (50 mm and 100 mm, 3 µm or 5 µm particles).
- UV detection at 280 nm.
Results and Discussion
- Salt Gradient Slope: Shallower slopes improved separation of low-abundance charge variants by extending retention and reducing co-elution with the main peak.
- Gradient Time: Extending gradient time from 5 to 20 minutes enhanced resolution but introduced peak broadening; an optimized slope/time balance is essential.
- Flow Rate: Within the tested range (0.4–0.8 mL/min), flow rate had negligible impact on resolution, allowing faster runs without sacrificing quality.
- Particle Size: Switching from 5 µm to 3 µm particles offered minor resolution gains but increased backpressure, limiting high-flow operation.
- Column Length: Doubling column length from 50 mm to 100 mm significantly increased variant resolution at the cost of longer retention times.
Practical Implications and Benefits
This optimized salt-gradient IEX method delivers a robust, high-resolution profiling of trastuzumab charge variants. It supports stringent QA/QC in biopharma development by providing clear separation of acidic and basic species, enabling accurate quantitation and monitoring of critical quality attributes.
Future Trends and Applications
Advances may include integration with mass spectrometric detection for structural characterization of variants, further miniaturization of columns to reduce solvent use, and automation of method scouting using machine learning to expedite method development across diverse mAbs.
Conclusion
The Shim-pack Bio IEX SP-NP columns coupled with a Nexera Bio UHPLC system present a flexible platform for salt-gradient charge variant analysis. Careful tuning of gradient slope, time, and column dimensions yields high-resolution separations, while flow rate and particle size can be adjusted to balance throughput and pressure constraints.
References
- Vlasak J., Ionescu R. Curr. Pharm. Biotechnol., 2008, 9(6), 468–481.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Charge variant analysis of mAb biotherapeutics by Nexera XS with a Shim-pack Bio IEX Column
2021|Shimadzu|Applications
Liquid Chromatograph NexeraTM XS Application News Charge variant analysis of mAb biotherapeutics by Nexera XS with a Shim-pack Bio IEX Column Ashutosh Shelar1 , Bhaumik Trivedi 1 , Navin Devadiga1, and Krutika Padhye 2 1 Shimadzu Analytical (India) Pvt. Ltd.,…
Key words
charge, chargevariants, variantscetuximab, cetuximabomalizumab, omalizumabmabs, mabsiex, iexrituximab, rituximabshim, shimbio, bionexera, nexeramain, mainpack, packtrastuzumab, trastuzumabvariant, variantcex
Development of Monoclonal Antibody Charge Variant Analysis Methods Using a BioResolve SCX mAb Column
2019|Waters|Applications
[ APPLICATION NOTE ] Development of Monoclonal Antibody Charge Variant Analysis Methods Using a BioResolve SCX mAb Column Hua Yang, Bill Warren, and Stephan Koza Waters Corporation, Milford, MA, USA APPLICATION BENEFITS INTRODUCTION Method development parameters for cation Charge heterogeneity…
Key words
bioresolve, bioresolvecharge, chargemab, mabvariant, variantscx, scxgradient, gradientminutes, minutesmonoclonal, monoclonalslope, slopeantibody, antibodysalt, saltcetuximab, cetuximabdevelopment, developmentcolumn, columnpeak
Agilent Biocolumns - Charge Variant Analysis - Application Compendium
2021|Agilent Technologies|Guides
Agilent Biocolumns Charge Variant Analysis Application Compendium Contents Background 2 Getting Started 3 How to Guide - Ion-Exchange Chromatography for Biomolecule Analysis - 5991-3775EN 4 Featured Application Notes 22 Convenient Customization of Your Cation Exchange Analysis 5994-3257EN 22 How…
Key words
return, returnsection, sectioncontents, contentsbuffer, buffercharge, chargeexchange, exchangevariants, variantsvariant, variantadvisor, advisorgradient, gradientmin, minmau, maugradients, gradientscation, cationcex
Charge Variant analysis of mAb Biosimilars by Nexera UHPLC with a Shim-pack Bio IEX Column
2019|Shimadzu|Applications
Application News Biopharma / Nexera UHPLC Charge Variant analysis of mAb Biosimilars by Nexera TM UHPLC with a Shim-pack TM Bio IEX Column No. AD-0195 Yonghai Lu and Zhaoqi Zhan Application Development & Support Centre, Shimadzu (Asia Pacific), Singapore …
Key words
bevacizumab, bevacizumabvariants, variantsbiosimilar, biosimilargradient, gradientarea, areacharge, chargesalt, saltnexera, nexeraiex, iexacidic, acidicvariant, variantshim, shimmain, mainpack, packbasic