SIMULATION OF A QUADRUPOLE MASS FILTER EMPLOYING A DIGITAL WAVEFORM AND DISCONTINUOUS ION INTRODUCTION TO OBTAIN HIGH RESOLUTION AND TRANSMISSION
Posters | 2019 | WatersInstrumentation
The performance of quadrupole mass filters is often constrained by an inverse trade-off between mass resolution and ion transmission. Achieving both high resolution and high transmission is critical for demanding analytical applications across research, industrial QA/QC, and environmental monitoring. This study introduces a novel approach combining a digital extended-coverage (EC) waveform with pulsed ion introduction to overcome this limitation.
The primary goal of this work is to theoretically demonstrate a quadrupole mass filter operating with discontinuous ion introduction and a digital EC waveform that yields resolutions exceeding 50,000 at more than 50% transmission. Computational simulations are employed to compare this EC approach with conventional harmonic waveforms under various stability regions and practical constraints.
A discontinuous scanning sequence pulses ions into the quadrupole during zero-voltage intervals of the EC waveform, then applies the driving waveform once ions are fully inside the field region. The inverse amplitude phase characteristic (iAPC) metric is used to quantify phase-dependent positional acceptance. Stability and peak simulations are conducted using a matrix method solution of the Mathieu/Hill equation, complemented by SIMION 3D v8.1 models incorporating realistic rod geometries and field imperfections.
Simulation tools and parameters:
Key findings include:
The proposed method offers substantial improvements in mass spectrometer selectivity and sensitivity by delivering high resolution without severe transmission loss. It is applicable to advanced proteomics, environmental trace analysis, and industrial QA/QC, where both separation power and ion throughput are essential.
Future developments may include:
This theoretical and computational investigation demonstrates that discontinuous ion introduction combined with a digital EC waveform can effectively decouple resolution and transmission. The approach achieves unprecedented performance metrics in quadrupole mass filters, paving the way for next-generation analytical instruments.
LC/MS, LC/MS/MS, LC/QQQ, LC/SQ
IndustriesManufacturerWaters
Summary
Significance of the topic
The performance of quadrupole mass filters is often constrained by an inverse trade-off between mass resolution and ion transmission. Achieving both high resolution and high transmission is critical for demanding analytical applications across research, industrial QA/QC, and environmental monitoring. This study introduces a novel approach combining a digital extended-coverage (EC) waveform with pulsed ion introduction to overcome this limitation.
Objectives and Overview of the Study
The primary goal of this work is to theoretically demonstrate a quadrupole mass filter operating with discontinuous ion introduction and a digital EC waveform that yields resolutions exceeding 50,000 at more than 50% transmission. Computational simulations are employed to compare this EC approach with conventional harmonic waveforms under various stability regions and practical constraints.
Methodology
A discontinuous scanning sequence pulses ions into the quadrupole during zero-voltage intervals of the EC waveform, then applies the driving waveform once ions are fully inside the field region. The inverse amplitude phase characteristic (iAPC) metric is used to quantify phase-dependent positional acceptance. Stability and peak simulations are conducted using a matrix method solution of the Mathieu/Hill equation, complemented by SIMION 3D v8.1 models incorporating realistic rod geometries and field imperfections.
Instrumentation Used
Simulation tools and parameters:
- Matrix method for Mathieu/Hill equation stability and peak shape simulation
- SIMION 3D v8.1 for modelling non-ideal rod geometries (truncated hyperbolic and round rods)
- Quadrupole parameters: rod radius 4 mm, length 130 mm, RF frequency 1 MHz, m/z 556, axial energy 0.2 eV
Main Results and Discussion
Key findings include:
- The EC waveform in the r1 and r12 stability tips provides extended regions of high acceptance across both x- and y-axes, largely independent of resolution.
- Theoretical 10% valley resolutions exceed 100,000 in the EC r12 region without hitting stability limits, and >50,000 at >50% transmission in r1.
- Transmission vs. resolution curves demonstrate a flat 100% transmission up to R~5,000, with limits of R~16,000 (c≈15) for r1 and over R~240,000 (c≈1) for r12 at 1 MHz.
- Practical factors such as rod shape (round vs. truncated hyperbolic), mechanical misalignment, and timing jitter were analysed, revealing that hyperbolic rods and tight tolerances preserve high performance, while timing jitter up to ±1 ps has minimal impact.
- Lower RF frequencies reduce voltage requirements (e.g., ±0.5/1.6 kV at 0.5 MHz) while maintaining resolutions around 60,000.
Benefits and Practical Applications
The proposed method offers substantial improvements in mass spectrometer selectivity and sensitivity by delivering high resolution without severe transmission loss. It is applicable to advanced proteomics, environmental trace analysis, and industrial QA/QC, where both separation power and ion throughput are essential.
Future Trends and Potential Applications
Future developments may include:
- Hardware implementation of high-voltage, high-frequency digital EC drivers
- Integration with TOF-based phase-space manipulation optics to further reduce velocity spread
- Miniaturised or portable quadrupole systems for field analysis
- Adaptation of EC scanning strategies for tandem MS and imaging applications
Conclusion
This theoretical and computational investigation demonstrates that discontinuous ion introduction combined with a digital EC waveform can effectively decouple resolution and transmission. The approach achieves unprecedented performance metrics in quadrupole mass filters, paving the way for next-generation analytical instruments.
Reference
- E. Sheretov, et al., Opportunities for optimization of the RF signal applied to electrodes of quadrupole mass spectrometers. Part II. EC signals. International Journal of Mass Spectrometry, 1987, Vol. 198, Issue 1, pp. 83–96.
- L. A. Pipes, Matrix Solution of Equations of the Mathieu-Hill Type. Journal of Applied Physics, 1953, 24, 902–910.
- SIMION 3D v8.1, Scientific Instrument Services Ltd.
- W. Paul, H. P. Reinghard, U. Von Zahn, Das elektrische Massenfilter als Massenspektrometer und Isotopentrenner. Zeitschrift für Physik, 1958, 152, 143–153.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
IN SEARCH OF THE PERFECT NOTCH: A NOVEL APPROACH TO THE OPTIMIZATION OF DIPOLE EXCITATION WAVEFORMS IN QUADRUPOLE MASS FILTERS
2023|Waters|Posters
IN SEARCH OF THE PERFECT NOTCH: A NOVEL APPROACH TO THE OPTIMIZATION OF DIPOLE EXCITATION WAVEFORMS IN QUADRUPOLE MASS FILTERS Authors: Keith Richardson, Martin Green, David Langridge Affiliations: Waters Corporation, Wilmslow, UK OVERVIEW The amplitudes and phases are then optimised…
Key words
dipole, dipolenotch, notchamplitudes, amplitudeswaveform, waveformwaveforms, waveformsexcitation, excitationfrequency, frequencyquadrupole, quadrupoletransmission, transmissiondipoles, dipolesphases, phasesangular, angularmight, mightexploration, explorationinvolve
High Capacity Electrostatic Ion Trap Mass Spectrometer and its Signal Processing Design and Simulation On increasing charge capacity of EITs There have been increasing attempts to make use of larger space for ion trapping and measurement of the image charge…
Key words
field, fieldcharge, chargeimage, imagefrequency, frequencyeit, eitfft, fftion, ionfocusing, focusingisochronous, isochronousmotion, motionpitch, pitchaxsim, axsimtransform, transformrotational, rotationalprecession
Simulation of the Effect of Mechanical Misalignment on High M/Z Resolution Cdms Trap Solutions
2025|Waters|Posters
SIMULATION OF THE EFFECT OF MECHANICAL MISALIGNMENT ON HIGH M/Z RESOLUTION CDMS TRAP SOLUTIONS David Langridge, Kevin Giles, Keith Richardson Waters Corporation, Wilmslow, UK OVERVIEW PURPOSE: Examination of the effect of realistic mechanical build tolerances on the performance of high…
Key words
trap, trapcdms, cdmsoffset, offsethalf, halfelectrode, electroderesolution, resolutioncase, caseassemblies, assembliesmechanical, mechanicaltolerances, tolerancesmonte, montecarlo, carlostability, stabilityconstructed, constructedsolutions
A GPU ACCELERATED CDMS TRAP TRAJECTORY SIMULATOR AND OPTIMISER
2023|Waters|Posters
A GPU ACCELERATED CDMS TRAP TRAJECTORY SIMULATOR AND OPTIMISER David Langridge, Keith Richardson, Jeffery Brown, Kevin Giles Waters Corporation, Wilmslow, UK INTRODUCTION Charge detection mass spectrometry (CDMS) is an increasingly popular method for the analysis of large and heterogeneous ions.…
Key words
trajectory, trajectoryelit, elitoptimiser, optimisertube, tubevoltages, voltagestrap, trapcdms, cdmssimulator, simulatorgeometries, geometriesoptimisation, optimisationgpu, gpuelectrode, electroderesolution, resolutiongeometry, geometryion