SPE Troubleshooting - Extra Peak

We, 28.1.2026
| Original article from: Phenomenex
An SPE technical tip focused on identifying the sources of interferences and extra peaks observed after sample extraction.
<p><strong>Phenomenex: </strong><span>SPE Troubleshooting - Extra Peak</span></p>

Phenomenex: SPE Troubleshooting - Extra Peak

Welcome to our new SPE technical tip!

What is the cause when you encounter interferences post-extraction regardless of the analytical or detection method? Possible sources of the mystery peak(s) are outlined below.

Possible Cause/Source   /   Suggested Remedy

  • Contaminated reagents: Analyze the purity of reagents and additives or try using these from a different supplier.
  • Sample pre-treatment: Compare chromatograms of pre-treated and non-pre-treated sample.
  • Mismatched injection solvent: Analytes that are dissolved in a strong injection solvent may retain poorly and thus appear as extra peaks. Using the initial mobile phase as the injection solvent is recommended.
  • Incorrect SPE method: The SPE sorbent might not have been properly conditioned and equilibrated. Typically, the strongest solvent used in the method should be used to condition the sorbent. Analyze the conditioning, loading and washing fractions to check if the same interferences appear.
  • SPE tube leaching: Before conditioning, clean the tube with elution solvent. If reagents or the nature of the sample are too harsh for the polypropylene SPE tube, a glass tube or PTFE tube could be used instead.
  • Degradation/side reaction: Check the reactivity and stability of all components in your sample. Degradation can happen if compounds are sensitive to factors such as heat, light, acids, bases, etc. Sample components may be reacting with each other or with reagents/additives.

We hope this tip was useful. See you soon for another tip!

Phenomenex
LinkedIn Logo
 

Related content

Two-Step Matrix Deposition for Improved MS Imaging Resolution

Posters
| 2026 | Shimadzu (ASMS)
Instrumentation
LC/MS, LC/MS/MS, LC/TOF, LC/HRMS, MALDI, MS Imaging
Manufacturer
Shimadzu
Industries
Pharma & Biopharma

Smart Monitoring & Optimization Solutions for SMB Processes: In-Line and Off-Line Tools for SMB Systems

Technical notes
| 2026 | KNAUER
Instrumentation
HPLC
Manufacturer
KNAUER
Industries
Pharma & Biopharma

Xevo MRT P10 Mass Spectrometer

Brochures and specifications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/TOF, LC/HRMS
Manufacturer
Waters
Industries
Metabolomics, Proteomics , Lipidomics

Large-Volume Dilute-and-Shoot Analysis of PFAS in Drinking Water Using the Altura PFAS Column

Applications
| 2026 | Agilent Technologies
Instrumentation
Consumables, LC columns, LC/MS, LC/MS/MS, LC/QQQ
Manufacturer
Agilent Technologies
Industries
Environmental

Hey, ho, oliGO – Comparison of ion-pairing systems for oligonucleotide analysis with HPLC-UV

Applications
| 2026 | KNAUER
Instrumentation
Consumables, LC columns, HPLC
Manufacturer
KNAUER
Industries
Pharma & Biopharma
 

Related articles

Considerations when working with solid samples and SPE
Article | Academy

Considerations when working with solid samples and SPE

Solid phase extraction enables clean-up, pre-concentration, and solvent switching. Proper solvent choice and SPE mode are key for effective solid sample preparation.
Phenomenex
tag
share
more
Why are my recoveries greater than 100%?
Article | Academy

Why are my recoveries greater than 100%?

Although 100% recovery is expected in SPE, values above 100% are sometimes observed. This article explains the most common causes and provides practical guidance on how to identify and resolve them.
Phenomenex
tag
share
more
SPE Troubleshooting: Understanding and Resolving Low Recovery in Solid Phase Extraction
Article | Academy

SPE Troubleshooting: Understanding and Resolving Low Recovery in Solid Phase Extraction

Low SPE recovery does not have to mean method failure. Learn how fraction collection and analysis can identify analyte losses and guide effective troubleshooting steps.
Phenomenex
tag
share
more
SPE Troubleshooting: Why Won't My Sample Flow Through the SPE Cartridge?
Article | Academy

SPE Troubleshooting: Why Won't My Sample Flow Through the SPE Cartridge?

Learn the most common causes of SPE cartridge flow problems and how to troubleshoot them. Identify issues related to samples, solvents, manifold pressure, and precipitation.
Phenomenex
tag
share
more
 

Related content

Two-Step Matrix Deposition for Improved MS Imaging Resolution

Posters
| 2026 | Shimadzu (ASMS)
Instrumentation
LC/MS, LC/MS/MS, LC/TOF, LC/HRMS, MALDI, MS Imaging
Manufacturer
Shimadzu
Industries
Pharma & Biopharma

Smart Monitoring & Optimization Solutions for SMB Processes: In-Line and Off-Line Tools for SMB Systems

Technical notes
| 2026 | KNAUER
Instrumentation
HPLC
Manufacturer
KNAUER
Industries
Pharma & Biopharma

Xevo MRT P10 Mass Spectrometer

Brochures and specifications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/TOF, LC/HRMS
Manufacturer
Waters
Industries
Metabolomics, Proteomics , Lipidomics

Large-Volume Dilute-and-Shoot Analysis of PFAS in Drinking Water Using the Altura PFAS Column

Applications
| 2026 | Agilent Technologies
Instrumentation
Consumables, LC columns, LC/MS, LC/MS/MS, LC/QQQ
Manufacturer
Agilent Technologies
Industries
Environmental

Hey, ho, oliGO – Comparison of ion-pairing systems for oligonucleotide analysis with HPLC-UV

Applications
| 2026 | KNAUER
Instrumentation
Consumables, LC columns, HPLC
Manufacturer
KNAUER
Industries
Pharma & Biopharma
 

Related articles

Considerations when working with solid samples and SPE
Article | Academy

Considerations when working with solid samples and SPE

Solid phase extraction enables clean-up, pre-concentration, and solvent switching. Proper solvent choice and SPE mode are key for effective solid sample preparation.
Phenomenex
tag
share
more
Why are my recoveries greater than 100%?
Article | Academy

Why are my recoveries greater than 100%?

Although 100% recovery is expected in SPE, values above 100% are sometimes observed. This article explains the most common causes and provides practical guidance on how to identify and resolve them.
Phenomenex
tag
share
more
SPE Troubleshooting: Understanding and Resolving Low Recovery in Solid Phase Extraction
Article | Academy

SPE Troubleshooting: Understanding and Resolving Low Recovery in Solid Phase Extraction

Low SPE recovery does not have to mean method failure. Learn how fraction collection and analysis can identify analyte losses and guide effective troubleshooting steps.
Phenomenex
tag
share
more
SPE Troubleshooting: Why Won't My Sample Flow Through the SPE Cartridge?
Article | Academy

SPE Troubleshooting: Why Won't My Sample Flow Through the SPE Cartridge?

Learn the most common causes of SPE cartridge flow problems and how to troubleshoot them. Identify issues related to samples, solvents, manifold pressure, and precipitation.
Phenomenex
tag
share
more
 

Related content

Two-Step Matrix Deposition for Improved MS Imaging Resolution

Posters
| 2026 | Shimadzu (ASMS)
Instrumentation
LC/MS, LC/MS/MS, LC/TOF, LC/HRMS, MALDI, MS Imaging
Manufacturer
Shimadzu
Industries
Pharma & Biopharma

Smart Monitoring & Optimization Solutions for SMB Processes: In-Line and Off-Line Tools for SMB Systems

Technical notes
| 2026 | KNAUER
Instrumentation
HPLC
Manufacturer
KNAUER
Industries
Pharma & Biopharma

Xevo MRT P10 Mass Spectrometer

Brochures and specifications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/TOF, LC/HRMS
Manufacturer
Waters
Industries
Metabolomics, Proteomics , Lipidomics

Large-Volume Dilute-and-Shoot Analysis of PFAS in Drinking Water Using the Altura PFAS Column

Applications
| 2026 | Agilent Technologies
Instrumentation
Consumables, LC columns, LC/MS, LC/MS/MS, LC/QQQ
Manufacturer
Agilent Technologies
Industries
Environmental

Hey, ho, oliGO – Comparison of ion-pairing systems for oligonucleotide analysis with HPLC-UV

Applications
| 2026 | KNAUER
Instrumentation
Consumables, LC columns, HPLC
Manufacturer
KNAUER
Industries
Pharma & Biopharma
 

Related articles

Considerations when working with solid samples and SPE
Article | Academy

Considerations when working with solid samples and SPE

Solid phase extraction enables clean-up, pre-concentration, and solvent switching. Proper solvent choice and SPE mode are key for effective solid sample preparation.
Phenomenex
tag
share
more
Why are my recoveries greater than 100%?
Article | Academy

Why are my recoveries greater than 100%?

Although 100% recovery is expected in SPE, values above 100% are sometimes observed. This article explains the most common causes and provides practical guidance on how to identify and resolve them.
Phenomenex
tag
share
more
SPE Troubleshooting: Understanding and Resolving Low Recovery in Solid Phase Extraction
Article | Academy

SPE Troubleshooting: Understanding and Resolving Low Recovery in Solid Phase Extraction

Low SPE recovery does not have to mean method failure. Learn how fraction collection and analysis can identify analyte losses and guide effective troubleshooting steps.
Phenomenex
tag
share
more
SPE Troubleshooting: Why Won't My Sample Flow Through the SPE Cartridge?
Article | Academy

SPE Troubleshooting: Why Won't My Sample Flow Through the SPE Cartridge?

Learn the most common causes of SPE cartridge flow problems and how to troubleshoot them. Identify issues related to samples, solvents, manifold pressure, and precipitation.
Phenomenex
tag
share
more
 

Related content

Two-Step Matrix Deposition for Improved MS Imaging Resolution

Posters
| 2026 | Shimadzu (ASMS)
Instrumentation
LC/MS, LC/MS/MS, LC/TOF, LC/HRMS, MALDI, MS Imaging
Manufacturer
Shimadzu
Industries
Pharma & Biopharma

Smart Monitoring & Optimization Solutions for SMB Processes: In-Line and Off-Line Tools for SMB Systems

Technical notes
| 2026 | KNAUER
Instrumentation
HPLC
Manufacturer
KNAUER
Industries
Pharma & Biopharma

Xevo MRT P10 Mass Spectrometer

Brochures and specifications
| 2026 | Waters
Instrumentation
LC/MS, LC/MS/MS, LC/TOF, LC/HRMS
Manufacturer
Waters
Industries
Metabolomics, Proteomics , Lipidomics

Large-Volume Dilute-and-Shoot Analysis of PFAS in Drinking Water Using the Altura PFAS Column

Applications
| 2026 | Agilent Technologies
Instrumentation
Consumables, LC columns, LC/MS, LC/MS/MS, LC/QQQ
Manufacturer
Agilent Technologies
Industries
Environmental

Hey, ho, oliGO – Comparison of ion-pairing systems for oligonucleotide analysis with HPLC-UV

Applications
| 2026 | KNAUER
Instrumentation
Consumables, LC columns, HPLC
Manufacturer
KNAUER
Industries
Pharma & Biopharma
 

Related articles

Considerations when working with solid samples and SPE
Article | Academy

Considerations when working with solid samples and SPE

Solid phase extraction enables clean-up, pre-concentration, and solvent switching. Proper solvent choice and SPE mode are key for effective solid sample preparation.
Phenomenex
tag
share
more
Why are my recoveries greater than 100%?
Article | Academy

Why are my recoveries greater than 100%?

Although 100% recovery is expected in SPE, values above 100% are sometimes observed. This article explains the most common causes and provides practical guidance on how to identify and resolve them.
Phenomenex
tag
share
more
SPE Troubleshooting: Understanding and Resolving Low Recovery in Solid Phase Extraction
Article | Academy

SPE Troubleshooting: Understanding and Resolving Low Recovery in Solid Phase Extraction

Low SPE recovery does not have to mean method failure. Learn how fraction collection and analysis can identify analyte losses and guide effective troubleshooting steps.
Phenomenex
tag
share
more
SPE Troubleshooting: Why Won't My Sample Flow Through the SPE Cartridge?
Article | Academy

SPE Troubleshooting: Why Won't My Sample Flow Through the SPE Cartridge?

Learn the most common causes of SPE cartridge flow problems and how to troubleshoot them. Identify issues related to samples, solvents, manifold pressure, and precipitation.
Phenomenex
tag
share
more
Other projects
GCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike