Plasmion SICRIT® Interface AX1/2 - Installation Manual
Manuals | 2020 | PlasmionInstrumentation
Flow-through plasma ionization offers a gentle and versatile approach to analyte detection in LC-MS workflows. By generating a cold plasma for ionization, SICRIT technology expands the range of detectable compounds and supports direct gas-phase analysis as well as coupling to chromatographic systems.
This manual provides detailed guidance for installing and configuring the SICRIT Interface AX1/2 on Agilent LC-MS instruments. It replaces the standard electrospray housing, enabling plasma-based ionization and modular integration with GC and SPME coupling modules.
Installation steps include removal of the original ESI source, assembly of a ceramic spacer and adaptor pieces, secure mounting of the SICRIT ion source with a quick lock mechanism, and installation of a safety interlock. The interlock not only protects users from high voltage but also ensures instrument recognition. Recommended software settings for Agilent MassHunter and ChemStation are provided, such as drying gas temperatures between 250 and 300 C, gas flows of 3 to 12 L min, zero nozzle voltage, and fragmentor voltages around 150 to 250 V.
The interface can be installed quickly without specialized tools, supports safe high-voltage integration, and maintains full compatibility with Agilent software. A described workaround for older instruments using the original iButton ensures seamless recognition. The design accommodates different capillary flows via distinct adaptor variants AX1 and AX2.
Advancements may include integration with automated sampling modules, expanded analyte libraries, miniaturized ion source designs, and coupling with high resolution and ion mobility spectrometers to enhance analytical throughput and selectivity.
The SICRIT Interface AX1/2 offers a user-friendly, safe, and effective solution for integrating plasma-based soft ionization into Agilent LC-MS systems, broadening analytical capabilities in research and quality control laboratories.
LC/MS
IndustriesManufacturerAgilent Technologies, Plasmion
Summary
Significance of the Topic
Flow-through plasma ionization offers a gentle and versatile approach to analyte detection in LC-MS workflows. By generating a cold plasma for ionization, SICRIT technology expands the range of detectable compounds and supports direct gas-phase analysis as well as coupling to chromatographic systems.
Objectives and Study Overview
This manual provides detailed guidance for installing and configuring the SICRIT Interface AX1/2 on Agilent LC-MS instruments. It replaces the standard electrospray housing, enabling plasma-based ionization and modular integration with GC and SPME coupling modules.
Methodology and Instrumentation
Installation steps include removal of the original ESI source, assembly of a ceramic spacer and adaptor pieces, secure mounting of the SICRIT ion source with a quick lock mechanism, and installation of a safety interlock. The interlock not only protects users from high voltage but also ensures instrument recognition. Recommended software settings for Agilent MassHunter and ChemStation are provided, such as drying gas temperatures between 250 and 300 C, gas flows of 3 to 12 L min, zero nozzle voltage, and fragmentor voltages around 150 to 250 V.
Main Results and Discussion
The interface can be installed quickly without specialized tools, supports safe high-voltage integration, and maintains full compatibility with Agilent software. A described workaround for older instruments using the original iButton ensures seamless recognition. The design accommodates different capillary flows via distinct adaptor variants AX1 and AX2.
Benefits and Practical Applications
- Soft ionization with minimal fragmentation increases analyte coverage.
- Direct retrofit to existing LC-MS instruments reduces downtime.
- Modular design supports GC and SPME coupling for diverse workflows.
- Interlock system provides user protection and maintains instrument readiness.
Future Trends and Potential Uses
Advancements may include integration with automated sampling modules, expanded analyte libraries, miniaturized ion source designs, and coupling with high resolution and ion mobility spectrometers to enhance analytical throughput and selectivity.
Conclusion
The SICRIT Interface AX1/2 offers a user-friendly, safe, and effective solution for integrating plasma-based soft ionization into Agilent LC-MS systems, broadening analytical capabilities in research and quality control laboratories.
Reference
- No references listed in the source document.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Plasmion SICRIT® Interface SX1 - Installation Manual
2020|Plasmion|Manuals
Installation Manual SICRIT® Interface SX1 for SCIEX LC-MS instruments Release January 2020 Plasmion GmbH – Am Mittleren Moos 48 - 86167 Augsburg - Germany Installation Manual SICRIT® Interface SX1 Version 0.2 This manual has to be to be stored carefully…
Key words
sicrit, sicritinterlock, interlocksource, sourceadaptor, adaptorsciex, sciexion, ioninterface, interfacediscs, discsspacer, spacerinstallation, installationmanual, manualventuri, venturidistance, distancedial, dialsure
Plasmion SICRIT® SC-30X Ionization Set - Hardware and Operations Manual
2020|Plasmion|Manuals
Hardware and Operations Manual SICRIT® SC-30X Ionization Set consisting of SICRIT® Control Unit SC-30 and SICRIT® Ion Source Release January 2020 Plasmion GmbH – Am Mittleren Moos 48 - 86167 Augsburg - Germany Hardware and Operations Manual SICRIT® SC-30X Ionization…
Key words
sicrit, sicritattention, attentionoutput, outputcard, cardoperation, operationstatus, statussource, sourceoff, offinterface, interfacedeactivate, deactivatevoltage, voltagemodule, moduleion, ionsafety, safetyslot
Plasmion SICRIT® Interface TX2/3 - Installation Manual
2020|Thermo Fisher Scientific|Manuals
Installation Manual SICRIT® Interface TX2/3 for Thermo Fisher LTQ, LTQ Velos, Orbitrap Elite and Exactive Series Release January 2020 Plasmion GmbH – Am Mittleren Moos 48 - 86167 Augsburg - Germany Installation Manual SICRIT® Interface TX2/3 Version 0.2 This manual…
Key words
sicrit, sicritinterlock, interlocksafety, safetycapillary, capillarymanual, manuallaunching, launchingattention, attentionrecommended, recommendedoperation, operationcap, capinlet, inletmanuals, manualsresponsibility, responsibilityinstallation, installationdismounted
Plasmion SICRIT® GC/SPME Module - Installation and Operations Manual
2020|Plasmion|Manuals
Installation and Operations Manual SICRIT® GC/SPME Module Release Jan 2020 Plasmion GmbH – Am Mittleren Moos 48 – 86167 Augsburg - Deutschland - Installation and Operations Manual SICRIT® GC/SPME-Module Version 0.3 This manual must be to be stored carefully and…
Key words
sicrit, sicritspme, spmemodule, moduleoperations, operationsinstallation, installationmanual, manualtransferline, transferlineplasmion, plasmionrail, railswagelok, swagelokliner, linerinsert, insertinlet, inlettight, tightcarefully