Agilent InfinityLab Quick Change Valve G5641A - Instructions
Manuals | 2021 | Agilent TechnologiesInstrumentation
Modern bioanalytical workflows demand rapid valve switching, efficient sample preparation and continuous column operation to maximize throughput and sensitivity. The InfinityLab Quick Change Bio Valve G5641A addresses these requirements by enabling flexible fluid routing, automated enrichment/cleanup and parallel column regeneration.
This technical note presents the installation and application of the Agilent InfinityLab Quick Change 2-position/10-port Bio Valve (G5641A) in a 1290 Infinity II Multicolumn Thermostat. The primary objectives are to demonstrate mechanical assembly, fluidic port mapping and practical use cases spanning dual-column selection, sample enrichment/cleanup and alternating column regeneration.
Installation steps:
Key instrumentation:
The G5641A valve supports three core workflows:
These configurations yield higher productivity, automated sample handling and reliable performance in demanding bioanalytical applications.
The InfinityLab Quick Change Bio Valve offers:
Modular design facilitates retrofits on existing 1290 platforms without special tools, boosting lab efficiency.
Emerging developments may include tighter integration with mass spectrometry for proteomics, expansion to multi-valve multiplexed LC configurations, and advanced software-driven valve scheduling. Advances in bioinert materials and microfluidic valve miniaturization are expected to further reduce carryover, lower dead volumes and enhance sensitivity in next-generation bioanalytical systems.
The Agilent InfinityLab Quick Change Bio Valve G5641A represents a versatile solution for high-throughput biochromatography. Its straightforward installation, robust fluidic design and flexible routing capabilities support diverse applications from automated sample prep to continuous gradient column regeneration, driving productivity and reliability in modern laboratories.
HPLC
IndustriesManufacturerAgilent Technologies
Summary
Significance of the Topic
Modern bioanalytical workflows demand rapid valve switching, efficient sample preparation and continuous column operation to maximize throughput and sensitivity. The InfinityLab Quick Change Bio Valve G5641A addresses these requirements by enabling flexible fluid routing, automated enrichment/cleanup and parallel column regeneration.
Goals and Study Overview
This technical note presents the installation and application of the Agilent InfinityLab Quick Change 2-position/10-port Bio Valve (G5641A) in a 1290 Infinity II Multicolumn Thermostat. The primary objectives are to demonstrate mechanical assembly, fluidic port mapping and practical use cases spanning dual-column selection, sample enrichment/cleanup and alternating column regeneration.
Methodology and Instrumentation
Installation steps:
- Remove transportation lock and dummy valve from the MCT rail.
- Mount the G5641A valve head onto the valve drive and secure with manual nut tightening.
- Attach Quick Connect Bio Heat Exchangers and MP35N capillaries, observing torque limits to avoid leaks.
- Map 6-port workflows onto the 10-port layout and install blank fittings on unused ports.
- Route capillaries through guides, stow excess tubing and perform leak checks.
Key instrumentation:
- Agilent 1290 Infinity II Multicolumn Thermostat (MCT)
- InfinityLab Quick Change Bio Valve G5641A (2-position/10-port, 1300 bar)
- InfinityLab Quick Connect Bio Heat Exchangers (Standard and High Flow)
- MP35N capillaries and InfinityLab Quick Turn/Quick Connect fittings
Main Results and Discussion
The G5641A valve supports three core workflows:
- Dual-column selection: instant switching between two analytical columns enhances separation flexibility and maintains uptime when a column ages.
- Sample enrichment and cleanup: on-line precolumn trapping with backflush strategies concentrates analytes or removes complex matrices, improving sensitivity and reproducibility.
- Alternating column regeneration: concurrent column equilibration during analysis on the alternate column can reduce idle regeneration time by up to 50 %, doubling sample throughput in gradient methods.
These configurations yield higher productivity, automated sample handling and reliable performance in demanding bioanalytical applications.
Benefits and Practical Applications
The InfinityLab Quick Change Bio Valve offers:
- Automated and reproducible sample enrichment/cleanup for complex matrices (e.g., food extracts, wastewater).
- Rapid method switching via dual-column selection for different stationary phases.
- High-pressure operation up to 1300 bar with bio-inert fluid paths (PEEK, MP35N).
- Integrated alternating regeneration to eliminate chromatography downtime.
Modular design facilitates retrofits on existing 1290 platforms without special tools, boosting lab efficiency.
Future Trends and Potential Applications
Emerging developments may include tighter integration with mass spectrometry for proteomics, expansion to multi-valve multiplexed LC configurations, and advanced software-driven valve scheduling. Advances in bioinert materials and microfluidic valve miniaturization are expected to further reduce carryover, lower dead volumes and enhance sensitivity in next-generation bioanalytical systems.
Conclusion
The Agilent InfinityLab Quick Change Bio Valve G5641A represents a versatile solution for high-throughput biochromatography. Its straightforward installation, robust fluidic design and flexible routing capabilities support diverse applications from automated sample prep to continuous gradient column regeneration, driving productivity and reliability in modern laboratories.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Agilent InfinityLab Quick Change Valve G5641A - Instructions
2020|Agilent Technologies|Manuals
Agilent InfinityLab Quick Change Valve G5641A Instructions Agilent G5641 A Valve Kit MCT - Instructions This technical note describes the installation and application of the Agilent InfinityLab Quick Change 2-position/10-port Bio Valve G5641A in a 1290 Infinity II Multicolumn Thermostat…
Key words
valve, valvepump, pumpstator, statorhead, headcau, cauexchanger, exchangerparts, partsregeneration, regenerationquick, quickfitting, fittingheat, heatcolumn, columnenrichment, enrichmentwaste, wasteconnect
Agilent InfinityLab Quick Change Valves G4231A/C and G4232C/D - Instructions
2018|Agilent Technologies|Manuals
Agilent InfinityLab Quick Change Valves G4231A/C and G4232C/D Instructions Quick Change Valves G4231A/C and G4232C/D This technical note describes the installation and application of the Agilent InfinityLab Quick Change 2ps/6pt Valve Heads G4231A/C and 2ps/10pt Valve Heads G4232C/D in a…
Key words
valve, valveexchanger, exchangerquick, quickheat, heatroxpq, roxpqfitting, fittingconnect, connectsxps, sxpsfroxpq, froxpqcolumn, columnstator, statorsocket, sockethead, headparts, partsturn
Agilent InfinityLab Quick Change Valve G5640A - Instructions
2022|Agilent Technologies|Manuals
Agilent InfinityLab Quick Change Valve G5640A Instructions This technical note describes the installation and application of the Agilent InfinityLab Quick Change 6-position/14-port Bio Valve G5640A in a 1290 Infinity II Multicolumn Thermostat (MCT). Typical Applications 2 Multicolumn Selection 2 Method…
Key words
valve, valveheads, headsparts, partsinstall, installhead, headstator, statortransportation, transportationexchanger, exchangertypical, typicalchecklist, checklistlock, locknote, notedummy, dummygroove, groovedamage
Agilent InfinityLab Quick Change Valve G5632A Instructions
2018|Agilent Technologies|Manuals
Agilent InfinityLab Quick Change Valve G5632A Instructions Agilent InfinityLab Quick Change Valve G5632A Instructions Contents Typical Applications 2 Alternating Column Regeneration (2 position/10 port valves only) 2 Dual column selection 3 Sample Enrichment and Sample Cleanup 4 Delivery Checklist Specifications…
Key words
valve, valvefroxpq, froxpqwlfdo, wlfdocapillaries, capillariescladded, claddedsxps, sxpshwhfwru, hwhfwruroxpq, roxpqqdo, qdoregeneration, regenerationcaution, cautionhead, headpeek, peekenrichment, enrichmentcapillary