Agilent InfinityLab Quick Change Valve G5640A - Instructions
Manuals | 2022 | Agilent TechnologiesInstrumentation
Fast and reliable selection of multiple chromatographic columns is critical in modern bio-LC workflows and method development. The Agilent InfinityLab Quick Change 6-position/14-port Bio Valve G5640A enables seamless switching among up to six columns under high pressure (1300 bar), minimizing downtime and accelerating experiment throughput.
This technical summary outlines the installation, configuration, and practical applications of the G5640A valve kit in the 1290 Infinity II Multicolumn Thermostat (MCT). It highlights key steps for hardware setup, capillary connections, and integration into routine and automated method development processes.
The installation procedure comprises:
Valve materials are optimized for bio applications, with PEEK and MP35N flow paths and compatibility across a pH range of 0–14 (except some mineral acids). Key instrument details:
Implementation of the G5640A valve delivers significant improvements in workflow efficiency and method flexibility. Users can:
Proper installation and use of bio-inert components prevent sample degradation and ensure reproducible, unbiased chromatographic performance.
The valve kit supports a range of applications:
The high pressure rating and robust construction ensure reliable operation under demanding conditions, maximizing instrument uptime.
Future developments may include integration with advanced control software for fully automated method optimization, miniaturized valve designs for micro-flow applications, and expanded compatibility with novel stationary phase materials. Cloud-based data analytics could leverage multicolumn datasets for AI-driven method recommendations.
The Agilent InfinityLab Quick Change G5640A valve in the 1290 Infinity II MCT presents a powerful solution for rapid, reliable column switching in bio-LC and method development. Its simple installation, bio-inert flow path, and high pressure tolerance translate into enhanced laboratory productivity and flexibility.
HPLC
IndustriesManufacturerAgilent Technologies
Summary
Importance of the Topic
Fast and reliable selection of multiple chromatographic columns is critical in modern bio-LC workflows and method development. The Agilent InfinityLab Quick Change 6-position/14-port Bio Valve G5640A enables seamless switching among up to six columns under high pressure (1300 bar), minimizing downtime and accelerating experiment throughput.
Objectives and Study Overview
This technical summary outlines the installation, configuration, and practical applications of the G5640A valve kit in the 1290 Infinity II Multicolumn Thermostat (MCT). It highlights key steps for hardware setup, capillary connections, and integration into routine and automated method development processes.
Methodology and Instrumentation
The installation procedure comprises:
- Removing the transportation lock and dummy valve from the MCT valve drive.
- Mounting the G5640A valve head by aligning guide pins with actuator grooves and hand-tightening the retaining nut.
- Connecting bio-compatible capillaries (MP35N, PEEK components) using M4 threaded fittings to all required ports (column, detector, sampler, bypass, waste).
- Power-cycling the module to allow RFID-based tag reader identification of valve properties.
Valve materials are optimized for bio applications, with PEEK and MP35N flow paths and compatibility across a pH range of 0–14 (except some mineral acids). Key instrument details:
- Module: Agilent 1290 Infinity II Multicolumn Thermostat (MCT)
- Valve head: Agilent InfinityLab Quick Change 6-position/14-port Bio Valve G5640A
- Capillary kit: Bio 6-Column Selector Capillary Kit (p/n 5005-0070) with MP35N capillaries, quick-connect fittings, heat exchangers and waste tubing.
Main Results and Discussion
Implementation of the G5640A valve delivers significant improvements in workflow efficiency and method flexibility. Users can:
- Rapidly switch among up to six stationary phases or column dimensions without manual reconnection.
- Shorten assay development time by parallelizing resolution, speed, and loading studies.
- Enable automated scouting sequences for optimal column selection in bio-molecule analysis.
Proper installation and use of bio-inert components prevent sample degradation and ensure reproducible, unbiased chromatographic performance.
Benefits and Practical Applications
The valve kit supports a range of applications:
- Multicolumn selection for high-throughput screening of stationary phases.
- Method development workflows requiring variable column lengths and internal diameters.
- Automated bio-LC analyses in pharmaceutical research, QA/QC, and proteomics.
The high pressure rating and robust construction ensure reliable operation under demanding conditions, maximizing instrument uptime.
Future Trends and Opportunities
Future developments may include integration with advanced control software for fully automated method optimization, miniaturized valve designs for micro-flow applications, and expanded compatibility with novel stationary phase materials. Cloud-based data analytics could leverage multicolumn datasets for AI-driven method recommendations.
Conclusion
The Agilent InfinityLab Quick Change G5640A valve in the 1290 Infinity II MCT presents a powerful solution for rapid, reliable column switching in bio-LC and method development. Its simple installation, bio-inert flow path, and high pressure tolerance translate into enhanced laboratory productivity and flexibility.
Used Instrumentation
- Agilent 1290 Infinity II Multicolumn Thermostat (MCT)
- InfinityLab Quick Change 6-position/14-port Bio Valve G5640A (p/n 5320-0025)
- Bio 6-Column Selector Capillary Kit (p/n 5005-0070) with MP35N capillaries
- Quick-Connect Bio Heat Exchanger fittings and related tools
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Agilent InfinityLab Quick Change Valve G5641A - Instructions
2021|Agilent Technologies|Manuals
Agilent InfinityLab Quick Change Valve G5641A Instructions This technical note describes the installation and application of the Agilent InfinityLab Quick Change 2-position/10-port Bio Valve G5641A in a 1290 Infinity II Multicolumn Thermostat (MCT). Typical Applications 2 Dual Column Selection (2-Position/6-Port…
Key words
valve, valveinstallation, installationstator, statorpump, pumpparts, partsquick, quickexchanger, exchangerfitting, fittingcolumn, columnhead, headconnect, connectenrichment, enrichmentwaste, wasteheat, heatcapillary
Agilent InfinityLab Quick Change Valve G5641A - Instructions
2020|Agilent Technologies|Manuals
Agilent InfinityLab Quick Change Valve G5641A Instructions Agilent G5641 A Valve Kit MCT - Instructions This technical note describes the installation and application of the Agilent InfinityLab Quick Change 2-position/10-port Bio Valve G5641A in a 1290 Infinity II Multicolumn Thermostat…
Key words
valve, valvepump, pumpstator, statorhead, headcau, cauexchanger, exchangerparts, partsregeneration, regenerationquick, quickfitting, fittingheat, heatcolumn, columnenrichment, enrichmentwaste, wasteconnect
Agilent InfinityLab Quick Change Valves G4234A/C - Instructions
2019|Agilent Technologies|Manuals
Agilent InfinityLab Quick Change Valves G4234A/C Instructions Agilent G4234A/C Valve Kit MCT - Instructions This technical note describes the installation and application of the Agilent InfinityLab Quick Change 6 Column Selector Valve Heads G4234A and G4234C in a 1290 Infinity…
Key words
valve, valvehead, headparts, partsstator, statorcaution, cautionbio, bioplq, plqinert, inertinstall, installcapillaries, capillariesnote, notetransportation, transportationdamage, damageexchanger, exchangerconnect
Agilent InfinityLab Quick Change Valve G4237A Instructions
2018|Agilent Technologies|Manuals
Agilent InfinityLab Quick Change Valve G4237A Instructions This technical note describes the installation and application of the Agilent InfinityLab Quick Change 4 column selector Valve Head G4237A in a 1260/1290 Infinity II Multicolumn Thermostat (MCT). Contents Typical Applications 2 Multicolumn…
Key words
valve, valvestator, statorhead, headfitting, fittingparts, partsexchanger, exchangercaution, cautionplq, plqnote, noteheat, heatring, ringinstall, installdamage, damagebio, biocolumn