Agilent InfinityLab Quick Change Valve G5641A - Instructions

Manuals | 2020 | Agilent TechnologiesInstrumentation
HPLC
Industries
Manufacturer
Agilent Technologies

Summary

Importance of the Topic


High-performance liquid chromatography workflows benefit from flexible, reliable valve solutions that support complex analytical methods such as dual column operation, sample cleanup, and alternating column regeneration. The Agilent InfinityLab Quick Change Bio Valve G5641A addresses these needs by offering high-pressure compatibility and rapid switching capabilities.

Objectives and Overview


This technical document outlines the installation procedure and application modes of the 2-position/10-port Bio Valve G5641A in the 1290 Infinity II Multicolumn Thermostat. It aims to guide users through delivery checks, system configuration, port mapping, and capillary routing to maximize instrument uptime and analytical throughput.

Methodology and Used Instrumentation


The workflow involves mounting the G5641A valve heads on Multicolumn Thermostats, installing Quick Connect heat exchangers, and routing capillaries using the supplied bio-compatible kits. Instrumentation highlights include:
  • 1290 Infinity II Multicolumn Thermostat (MCT)
  • 2-position/10-port Bio Valve G5641A (up to 1300 bar)
  • Quick Connect and Quick Turn MP35N capillaries
  • Bio-compatible PEEK and MP35N fittings

Main Results and Discussion


The valve supports several operational modes:
  • Dual column selection allows switching between stationary phases or using a backup column with minimal delay.
  • Sample enrichment and cleanup enable large-volume injections and matrix removal through precolumn trapping with backflushing.
  • Alternating column regeneration achieves parallel analysis and column equilibration, reducing idle time by up to 50 %.
Port mapping tables and installation checklists ensure correct routing without dead volumes or leaks.

Benefits and Practical Applications


Key advantages include increased productivity, improved reproducibility, and enhanced sensitivity for applications in proteomics, drug metabolism, environmental trace analysis, and complex matrix cleanup.

Future Trends and Opportunities


Emerging needs for higher throughput and automation suggest further integration of smart valve control within LC-MS systems, on-column workflows, and real-time method optimization. Miniaturized and multiplexed valve designs could expand applications in microflow and nanoflow analytics.

Conclusion


The Agilent InfinityLab Quick Change Bio Valve G5641A offers a robust solution for versatile high-pressure valve requirements in advanced liquid chromatography methods. Its flexible configurations streamline complex workflows, boost instrument utilization, and support diverse analytical challenges.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Agilent InfinityLab Quick Change Valve G5641A - Instructions
Agilent InfinityLab Quick Change Valve G5641A Instructions This technical note describes the installation and application of the Agilent InfinityLab Quick Change 2-position/10-port Bio Valve G5641A in a 1290 Infinity II Multicolumn Thermostat (MCT). Typical Applications 2 Dual Column Selection (2-Position/6-Port…
Key words
valve, valveinstallation, installationstator, statorpump, pumpparts, partsquick, quickexchanger, exchangerfitting, fittingcolumn, columnhead, headconnect, connectenrichment, enrichmentwaste, wasteheat, heatcapillary
Agilent InfinityLab Quick Change Valves G4231A/C and G4232C/D - Instructions
Agilent InfinityLab Quick Change Valves G4231A/C and G4232C/D Instructions Quick Change Valves G4231A/C and G4232C/D This technical note describes the installation and application of the Agilent InfinityLab Quick Change 2ps/6pt Valve Heads G4231A/C and 2ps/10pt Valve Heads G4232C/D in a…
Key words
valve, valveexchanger, exchangerquick, quickheat, heatroxpq, roxpqfitting, fittingsxps, sxpsconnect, connectfroxpq, froxpqcolumn, columnstator, statorsocket, sockethead, headparts, partsturn
Agilent InfinityLab Quick Change Valve G5640A - Instructions
Agilent InfinityLab Quick Change Valve G5640A Instructions This technical note describes the installation and application of the Agilent InfinityLab Quick Change 6-position/14-port Bio Valve G5640A in a 1290 Infinity II Multicolumn Thermostat (MCT). Typical Applications 2 Multicolumn Selection 2 Method…
Key words
valve, valveheads, headsparts, partsinstall, installhead, headstator, statortransportation, transportationexchanger, exchangertypical, typicalchecklist, checklistlock, locknote, notedummy, dummygroove, groovedamage
Agilent InfinityLab Quick Change Valve G5632A Instructions
Agilent InfinityLab Quick Change Valve G5632A Instructions Agilent InfinityLab Quick Change Valve G5632A Instructions Contents Typical Applications 2 Alternating Column Regeneration (2 position/10 port valves only) 2 Dual column selection 3 Sample Enrichment and Sample Cleanup 4 Delivery Checklist Specifications…
Key words
valve, valvefroxpq, froxpqwlfdo, wlfdosxps, sxpscladded, claddedcapillaries, capillarieshwhfwru, hwhfwruroxpq, roxpqqdo, qdoregeneration, regenerationcaution, cautionhead, headpeek, peekenrichment, enrichmentcapillary
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike