From HPLC Instrument to Column - Tracking Down the HPLC Problem
Presentations | 2022 | Agilent TechnologiesInstrumentation
The reliability of liquid chromatography is crucial for ensuring accurate analytical results in research and QA/QC environments. Systematic troubleshooting minimizes downtime and maintains method performance.
This guide offers a structured approach to identify sources of common LC issues from the pump to the column and detector. It categorizes problems into system pressure, peak shape, separation, and detection challenges and provides logical elimination steps.
The approach leverages Agilent Lab Advisor software for routine diagnostics and maintenance feedback. Key troubleshooting steps include pressure profiling via bypass valve switching, component isolation by sequentially disconnecting fittings, and UV lamp performance assessments to monitor baseline noise and drift.
System pressure issues such as blockages, leaks, and fluctuations are resolved by logical isolation of flow path components (needle seat, purge valves, capillaries) and routine preventive maintenance. Peak shape anomalies—including tailing, broadening, fronting, and splitting—are addressed through column cleaning or replacement, injection solvent optimization, and ensuring minimal dead volume with proper fittings. Retention time shifts and ghost peaks are controlled by consistent mobile phase preparation, buffer stability, and regular glassware cleaning. Detection stability is maintained through UV lamp intensity, noise and drift tests, and flow cell inspections.
Implementing this systematic troubleshooting framework extends column lifetime, enhances run-to-run reproducibility, and reduces overall downtime. Laboratories benefit from faster root-cause identification, leading to improved data quality and operational efficiency.
Emerging directions include AI-driven diagnostics for predictive maintenance, automated mobile phase management to prevent contamination, advanced column chemistries for higher robustness, and remote monitoring platforms to further streamline instrument uptime and support.
A structured, stepwise approach from instrument to column combines preventive maintenance, logical fault isolation, and regular performance checks to ensure consistent LC performance and minimize analytical disruptions.
No formal literature references provided in the original document.
Consumables, HPLC
IndustriesManufacturerAgilent Technologies
Summary
Significance of the Topic
The reliability of liquid chromatography is crucial for ensuring accurate analytical results in research and QA/QC environments. Systematic troubleshooting minimizes downtime and maintains method performance.
Goals and Study Overview
This guide offers a structured approach to identify sources of common LC issues from the pump to the column and detector. It categorizes problems into system pressure, peak shape, separation, and detection challenges and provides logical elimination steps.
Methodology and Instrumentation
The approach leverages Agilent Lab Advisor software for routine diagnostics and maintenance feedback. Key troubleshooting steps include pressure profiling via bypass valve switching, component isolation by sequentially disconnecting fittings, and UV lamp performance assessments to monitor baseline noise and drift.
Used Instrumentation
- Agilent 1260 and 1290 Infinity II LC systems (autosampler, pump, column compartment, UV diode array detector)
- Agilent Lab Advisor for calibration, diagnosis, and early maintenance feedback
- InfinityLab Quick Connect and Quick Turn fittings, Stay Safe caps for solvent bottles
- InfinityLab Poroshell 120 columns and Agilent Captiva syringe filters
Main Results and Discussion
System pressure issues such as blockages, leaks, and fluctuations are resolved by logical isolation of flow path components (needle seat, purge valves, capillaries) and routine preventive maintenance. Peak shape anomalies—including tailing, broadening, fronting, and splitting—are addressed through column cleaning or replacement, injection solvent optimization, and ensuring minimal dead volume with proper fittings. Retention time shifts and ghost peaks are controlled by consistent mobile phase preparation, buffer stability, and regular glassware cleaning. Detection stability is maintained through UV lamp intensity, noise and drift tests, and flow cell inspections.
Benefits and Practical Applications
Implementing this systematic troubleshooting framework extends column lifetime, enhances run-to-run reproducibility, and reduces overall downtime. Laboratories benefit from faster root-cause identification, leading to improved data quality and operational efficiency.
Future Trends and Applications
Emerging directions include AI-driven diagnostics for predictive maintenance, automated mobile phase management to prevent contamination, advanced column chemistries for higher robustness, and remote monitoring platforms to further streamline instrument uptime and support.
Conclusion
A structured, stepwise approach from instrument to column combines preventive maintenance, logical fault isolation, and regular performance checks to ensure consistent LC performance and minimize analytical disruptions.
Reference
No formal literature references provided in the original document.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
It Isn’t Always the Column: Troubleshooting Your HPLC Separation
2023|Agilent Technologies|Presentations
It Isn’t Always the Column: Troubleshooting Your HPLC Separation Jean Lane Applications Engineer LC Columns and Consumables Technical Support July 25, 2023 1 July 24, 2023 It Isn't Always The Column: Troubleshooting Your HPLC Separation DE25843549 You’ve Recognized that There…
Key words
mobile, mobilephase, phasecolumn, columnaction, actioncheck, checkpeak, peakpressure, pressuresolvent, solventshape, shapereplace, replaceuse, usechanges, changesyour, yourfitting, fittingparts
My Chromatography Has Changed: Steps for Effective Troubleshooting
2024|Agilent Technologies|Presentations
My Chromatography Has Changed: Steps for Effective Troubleshooting Jean Lane Applications Engineer LC Columns and Consumables Technical Support June 18, 2024 1 6/18/2024 My Chromatography has Changed: Steps for Effective Troubleshooting DE13277529 Common Symptoms and Problems • Increased pressure •…
Key words
mobile, mobilephase, phasecolumn, columnfilter, filterpressure, pressurecheck, checkaction, actionfitting, fittingpeak, peakquick, quickyour, yourconnections, connectionsuse, useinfinitylab, infinitylabreplace
Why Did That Happen?
2021|Agilent Technologies|Presentations
“"Why Did That Happen?"” Golnar Javadi Applications Engineer LC Columns and Consumables Technical Support March 31, 2021 "Why Did That Happen?" Common chromatography problems Changes in System Pressure • High pressure • Low pressure • Pressure fluctuations 1 Changes in…
Key words
happen, happendid, didwhy, whyporoshell, poroshellfilter, filteryes, yeselut, elutpressure, pressureinfinitylab, infinitylabchanges, changesspe, spesolvent, solventphase, phaseinline, inlinecaptiva
Surviving Chromatography - Part 2: Corrective Action
2018|Agilent Technologies|Presentations
Surviving Chromatography Part 2: Corrective Action Jean Lane Applications Engineer LC Columns & Consumables Technical Support October 11, 2018 1 October 10, 2018 Agilent restricted “My Chromatography is Not Where it Should be” – Points for Discussion for chromatography that…
Key words
column, columnplugged, pluggedpressure, pressurefrit, fritmobile, mobilereplace, replacephase, phaseshape, shapeproblems, problemscauses, causessolvent, solventpeak, peakpeaks, peaksdetach, detachecv