LCMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

Tips to Help Maximize Resolution

Presentations | 2022 | Agilent TechnologiesInstrumentation
HPLC
Industries
Manufacturer
Agilent Technologies

Summary

Importance of the Topic


In reversed-phase high-performance liquid chromatography, resolution determines the capacity to distinguish closely eluting compounds. Optimizing resolution is essential for reliable quantification, improved sensitivity, and robust method transfer in pharmaceutical, environmental, and biochemical analyses.

Study Objectives and Overview


This guide presents a structured method development plan focused on maximizing resolution by selecting appropriate column particle size and chemistry, fine-tuning mobile phase composition, and optimizing instrument configuration to achieve high efficiency and throughput.

Methodology and Instrumentation


  • Particle selection: superficially porous particles (SPP) at 1.9, 2.7, and 4 μm for fast separations with high plate counts.
  • Bonded phases: EC-C18, SB-C18, HPH-C18, CS-C18, phenyl-hexyl, and specialized chemistries to tailor selectivity via hydrophobic, π–π, and ionic interactions.
  • Mobile phases: binary gradients using aqueous buffers (formic, acetic, phosphate, carbonate, ammonium formate) and organic modifiers (acetonitrile, methanol); pH adjustments to modulate retention of acidic and basic analytes.
  • Instrumentation: Agilent InfinityLab LC systems (1100, 1260, 1290) with binary/quaternary pumps, temperature-controlled columns, DAD/VWD detectors, and optional LC/MS (ESI+, dMRM).
  • Accessories: Quick Connect/Quick Turn fittings, optimized tubing (0.12 mm ID, minimized length) to reduce extracolumn volume and dispersion.
  • Autosampler and detector settings: draw position, bottom sensing, data rates ≥80 Hz, bandwidth and reference wavelength selection for optimal peak definition.

Main Results and Discussion


  • Switching from 5 μm fully porous to 2.7 μm SPP columns reduced run times from ~35 min to <10 min while maintaining Rs ≥2 for critical peak pairs.
  • SPP columns deliver up to 200% higher efficiency than larger particles at moderate pressures (<600 bar).
  • Comparative evaluation of EC-C18, SB-C18, HPH-C18, and CS-C18 phases demonstrated distinct selectivity profiles, enabling baseline separation of diverse analyte mixtures.
  • Mobile phase pH was shown to directly influence retention: low pH enhances acid retention, high pH enhances basic compound retention.
  • Minimizing gradient dwell volume and extracolumn dispersion sharpened early-eluting peaks and accelerated system equilibration.
  • Robustness testing across multiple column lots with ΔRs monitoring ensured method reproducibility and transferability.

Benefits and Practical Applications


  • A systematic framework for method development reduces trial-and-error, accelerating optimization.
  • Enhanced throughput for QC/QA laboratories without sacrificing resolution or sensitivity.
  • Flexibility to analyze a broad range of analytes—from pharmaceuticals to environmental contaminants—by leveraging varied column chemistries and buffer pH adjustments.
  • Lower solvent consumption achieved by using narrow-bore columns and high-efficiency SPP technology.

Future Trends and Opportunities


Advances in superficially porous particle design aim to further bridge the gap to UHPLC performance at lower pressures. Automated dwell volume compensation and multi-phase method scouting will streamline development. Emerging multimodal and bioinert stationary phases will extend applications to proteomics and complex biopharmaceutical separations.

Conclusion


Maximizing resolution in LC requires an integrated approach: choosing the right particle size and bonded phase, fine-tuning mobile phase composition, and minimizing system dispersion. This methodology delivers robust, fast, and reproducible separations across diverse analytical challenges.

Reference


  • Agilent application note 5990-5572EN
  • Agilent application note 5994-2358EN
  • Agilent application note 5994-2274EN
  • Agilent application note 5988-6476EN
  • Agilent application note 5991-7560EN

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
What Do I Need for Successful Method Development?
What Do I Need for Successful Method Development?
2020|Agilent Technologies|Presentations
What Do I Need for Successful Method Development? Mark Powell Columns and Supplies Technical Support December 10, 2020 What is My Method Development Plan? 1. Smaller particles and superficially porous particles offer fast, efficient analysis 2. C18 column – general-purpose…
Key words
what, whatsuccessful, successfuldevelopment, developmentneed, needmethod, methodyes, yesmau, mauinfinitylab, infinitylabmobile, mobileabsorbance, absorbancephase, phasechoose, choosebonded, bondedvolume, volumecolumn
Getting Off to a Good Start - Isocratic method development
Getting Off to a Good Start - Isocratic method development
2020|Agilent Technologies|Presentations
Getting Off to a Good Start Isocratic method development Paul Altiero Applications Chemist, CSD February 13, 2020 1 February 13. 2020 Getting Off to a Good Start For Research Use Only. Not for use in diagnostic procedures. DE322025463 Isocratic Method…
Key words
getting, gettingporoshell, poroshellstart, startoff, offgood, goodinfinitylab, infinitylabbonded, bondedscouting, scoutingcolumn, columnwhat, whatmau, mauselectivity, selectivityisocratic, isocraticimpacts, impactsphase
How Do I Choose? A guide to HPLC column selection
How Do I Choose? A guide to HPLC column selection
2021|Agilent Technologies|Presentations
How Do I Choose? A guide to HPLC column selection Mark Powell Columns and Supplies Technical Support 16 December 2021 Overview How do I choose? • Efficiency • Particle size • Column length • Selectivity • Bonded phase • Mobile…
Key words
poroshell, poroshellchoose, choosehow, howhilic, hilicselectivity, selectivityuhplc, uhplcinfinitylab, infinitylabphase, phasemobile, mobilepolar, polarmau, maualternative, alternativechiral, chiralzorbax, zorbaxchemistries
Updating Old Methods:Is the Gain Worth the Pain? - Becoming a Better Chromatographer HPLC educational webinar
Updating Old Methods: Is the Gain Worth the Pain? Becoming a Better Chromatographer HPLC educational webinar Mark Powell Applications Engineer Columns and Supplies Technical Support Why Would You Consider Updating? What Is the Goal? • Do I need to update?…
Key words
mau, mauchiral, chiralmin, mininfinitylab, infinitylabporoshell, poroshellparticle, particlecolumn, columnagilent, agilentdeemter, deemterallowed, alloweddiffusion, diffusionvan, vanphase, phasechanges, changesmobile
Other projects
GCMS
ICPMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike