Proper Implementation of Enhanced Mechanical Calibration of Dissolution Apparatus 1 and 2

Technical notes | 2012 | Agilent TechnologiesInstrumentation
Dissolution
Industries
Pharma & Biopharma
Manufacturer
Agilent Technologies

Summary

Význam tématu


Enhanced mechanical calibration supports consistent dissolution results, optimizes instrument performance, and aligns with regulatory guidance to safeguard product quality.

Cíle a přehled studie / článku


This white paper presents a comprehensive approach to implement FDA and ASTM recommendations for enhanced mechanical qualification of dissolution apparatus, covering five key elements necessary for a complete calibration program.

Použitá metodika a instrumentace


The methodology comprises five components:
  • Enhanced mechanical measurements: RPM accuracy, vessel centering, wobble, and verticality;
  • Verification of dissolution components: dimensional checks of paddle, basket, shafts, and vessels against USP specifications;
  • Preventive maintenance: scheduled inspections and lubrication of mechanical parts;
  • Control of vessel variability, vibration, and dissolved gases: use of consistent vessels, vibration isolation, and thorough degassing protocols;
  • Prerun analyst checks: visual and functional inspection of all components before each test.
Instrumention includes Dissolution Apparatus 1 and 2, NIST-traceable gauges, run-out indicators, height and centering tools (e.g., Agilent 5010), vacuum degassing units, and vibration-dampening pads.

Hlavní výsledky a diskuse


Laboratories often focus on mechanical measurements but neglect complementary requirements such as maintenance, component verification and prerun checks, leading to incomplete qualification. Certificates of conformance must include individual measurements with traceable tools. Proper vessel and basket inspection reduces variability, while controlled degassing and vibration isolation preserve test integrity.

Přínosy a praktické využití metody


Adopting a full mechanical calibration workflow improves test reproducibility, shortens qualification time and cost compared to pharmacopoeial methods, ensures cGMP compliance, and supports data integrity in routine and troubleshooting scenarios.

Budoucí trendy a možnosti využití


Future developments include establishing quantitative vibration specifications, automated and digital calibration platforms, enhanced profiling of vessel surfaces, and integration with laboratory information management systems for real-time quality monitoring.

Závěr


When all five elements of enhanced mechanical calibration are implemented and documented, dissolution apparatus performance is optimized, variability is minimized, and regulatory expectations are met, leading to more reliable and efficient testing.

Reference


  1. PhRMA Subcommittee on Dissolution Calibration. Pharmacopoeial Forum, 2000;26(4):1149-1151.
  2. FDA Guidance for Industry. Mechanical Calibration of Dissolution Apparatus 1 and 2, 2010.
  3. ASTM E2503-07 Standard Practice for Qualification of Dissolution Apparatus, 2007.
  4. Moore TW et al. Implementation Guidance for ASTM E2503-07. Open Drug Delivery J., 2010;4:14-20.
  5. Gao Z et al. Gauge Repeatability and Reproducibility in Dissolution Testing. AAPS PharmSciTech, 2007;8(4):E1-E5.
  6. Salt A, Glennon J. Enhanced Mechanical Calibration of Dissolution Equipment. Dissolution Technologies, 2011.
  7. USP 35-NF30 <711> Dissolution, 2012.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Agilent 280-DS Mechanical Qualification System
Agilent 280-DS Mechanical Qualification System
2012|Agilent Technologies|Brochures and specifications
Agilent 280-DS Mechanical Qualification System REDEFINE DISSOLUTION QUALIFICATION SIMPLIFY MECHANICAL QUALIFICATION Maximize Laboratory Efficiency and Data Integrity The Agilent 280-DS Mechanical Qualification System (MQS) will change your perspective on dissolution qualification. Using the 280-DS’s two devices – an Instrument Module…
Key words
qualification, qualificationdissolution, dissolutionvessel, vesselapparatus, apparatusvibration, vibrationpvt, pvtyour, youragilent, agilentperiodic, periodicall, allcompliance, complianceservices, servicesbasket, basketedition, editionusp
Agilent 280-DS Mechanical Qualification System for Dissolution Apparatus 1 and 2
Agilent 280-DS Mechanical Qualification System for Dissolution Apparatus 1 and 2 A User-friendly System with Software Built for 21 CFR part 11 Compliance Technical Overview Authors Introduction Vinayak A.K and Dan Spisak The debate for the periodic qualification procedure of…
Key words
user, uservessel, vesselrecords, recordsqualification, qualificationbasket, basketmechanical, mechanicaldissolution, dissolutionsensors, sensorselectronic, electronicastm, astmfda, fdawobble, wobblereview, reviewcan, canfeatures
Mechanical Qualification Guide
Mechanical Qualification Guide
2012|Agilent Technologies|Posters
Mechanical Qualification Guide Vessel Measurement Dissolution Mechanical Qualification Specifications and Tolerances ICH Harmonized (USP, JP, EP) FDA DPA-LOP.002 ASTM E2503-07 Basket and paddle depth 25 ± 2 mm 25 ± 2 mm 25 ± 2 mm (or <8%) 23-27 mm…
Key words
basket, basketvessel, vesselwobble, wobblepaddle, paddleshaft, shaftverticality, verticalityrunout, runoutstainless, stainlesscentering, centeringsteel, steelmeasured, measureddissolution, dissolutiondamaged, damagedexamination, examinationsensors
AGILENT 708-DS DISSOLUTION APPARATUS TRAINING MANUAL
Agilent 708-DS Dissolution Apparatus Training Manual AGILENT 708-DS DISSOLUTION APPARATUS TRAINING MANUAL Table of Contents 1.1. The Apparatus .......................................................................................................................... - 5 - 1.2. The Firmware ............................................................................................................................ - 6 - 1.3. Optional Configurations ........................................................................................................ - 6 - 2. INSTRUMENT SETUP .............................................................................…
Key words
revision, revisiontraining, trainingmanual, manualanywhere, anywherebox, boxpress, pressvessel, vesselspindle, spindleinside, insideshaft, shaftbath, bathmanifold, manifoldddm, ddmenter, enterscreen
Other projects
GCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike