A TMTpro 18plex Proteomics Standard for Assessing Protein Measurement Accuracy and Precision
Posters | 2021 | Thermo Fisher ScientificInstrumentation
The development of robust multiplexed quantitation tools is critical for modern proteomics workflows, allowing simultaneous measurement of protein levels across many samples in a single LC-MS run. Accurate and precise assessment of reporter-ion–based methods such as Tandem Mass Tags (TMT) is essential for applications ranging from quantitative biology to quality control in bioprocessing. A standardized yeast digest labeled with TMTpro 18plex reagents provides a practical reference to benchmark instrument performance, optimize acquisition methods, and ensure consistent data quality across labs.
This study presents a prototype 18-plex TMTpro yeast digest standard designed to evaluate the accuracy and precision of protein quantitation using different LC-MS acquisition strategies. By mixing one parental yeast strain (BY4741) with three single-gene knockout strains (Met6, His4, Ura2) at defined ratios, the authors generated linear abundance series to test data-dependent MS2, synchronous precursor selection (SPS) MS3, and real-time search (RTS) MS3 methods on orbitrap-based platforms. The primary objectives were to compare quantitation accuracy, depth of coverage, and interference suppression across these workflows.
Sample Preparation
Data Acquisition
Data Processing
The following systems were employed:
Quantitation Accuracy and Interference
RTS-MS3 Performance
Comparative Depth
This TMTpro 18plex standard serves as a versatile benchmark for:
Expanding on this work, next-generation standards could:
The prototype TMTpro 18plex yeast digest standard effectively benchmarks quantitative accuracy and precision across MS2, SPS-MS3, and RTS-MS3 workflows. RTS-MS3 methods consistently yield higher proteome coverage and improved measurement fidelity, particularly for low-abundance channels. Adoption of this standard enables rigorous method validation, supports reproducible multi-site studies, and accelerates innovation in multiplexed proteomics.
Consumables, LC/HRMS, LC/MS, LC/MS/MS, LC/Orbitrap
IndustriesProteomics
ManufacturerThermo Fisher Scientific
Summary
Importance of the Topic
The development of robust multiplexed quantitation tools is critical for modern proteomics workflows, allowing simultaneous measurement of protein levels across many samples in a single LC-MS run. Accurate and precise assessment of reporter-ion–based methods such as Tandem Mass Tags (TMT) is essential for applications ranging from quantitative biology to quality control in bioprocessing. A standardized yeast digest labeled with TMTpro 18plex reagents provides a practical reference to benchmark instrument performance, optimize acquisition methods, and ensure consistent data quality across labs.
Goals and Overview
This study presents a prototype 18-plex TMTpro yeast digest standard designed to evaluate the accuracy and precision of protein quantitation using different LC-MS acquisition strategies. By mixing one parental yeast strain (BY4741) with three single-gene knockout strains (Met6, His4, Ura2) at defined ratios, the authors generated linear abundance series to test data-dependent MS2, synchronous precursor selection (SPS) MS3, and real-time search (RTS) MS3 methods on orbitrap-based platforms. The primary objectives were to compare quantitation accuracy, depth of coverage, and interference suppression across these workflows.
Methodology
Sample Preparation
- Four Saccharomyces cerevisiae strains (wild type and three knockouts) grown to mid-log phase in YPD.
- Cell lysis by bead beating, reduction, alkylation, and digestion with LysC and trypsin.
- Individual peptides labeled with unique TMTpro 18plex reagents and combined in triplicate to yield abundance points at 100%, 50%, 25%, 12.5%, 6.25%, and 0% for each knockout protein.
- C18 solid-phase cleanup prior to LC-MS analysis.
Data Acquisition
- Gradient elution (50 or 120 min) on nanoflow UHPLC with EASY-Spray C18 columns.
- Acquisition modes compared: high-resolution MS2, SPS-MS3, RTS-MS3 (with various close-out parameters selecting 5 or 10 peptides per trigger).
- Use of a FAIMS Pro interface assessed for its impact on interference reduction.
Data Processing
- Proteome Discoverer 2.5/3.0 with SEQUEST HT and COMET search engines.
- TMTpro mass tag defined as dynamic modification; 1% false discovery rate enforced at peptide level.
- Co-isolation thresholds set to 50% for MS2 and 70% for SPS-MS3 quantitation.
Instrumentation
The following systems were employed:
- Thermo Scientific Orbitrap Eclipse Tribrid mass spectrometer
- Thermo Scientific Exploris 480 mass spectrometer
- FAIMS Pro interface (optional)
- Thermo Scientific EASY-Spray ion source and C18 columns
- UltiMate 3000 RSLCnano and EASY-nLC 1200 UHPLC systems
Results and Discussion
Quantitation Accuracy and Interference
- SPS-MS3 provided superior accuracy and precision compared to MS2 methods, reducing ratio compression from co-isolated signals.
- The addition of FAIMS Pro further improved measurement accuracy for both MS2 and MS3 workflows by filtering out background ions.
RTS-MS3 Performance
- Real-time search (RTS) enhanced peptide identification rates, yielding ~26% more quantifiable proteins than SPS-MS3.
- Linearity of measured vs. expected protein ratios exceeded R2 = 0.98 for the first four dilution points; slight deviation (R2 ≈ 0.90) at the lowest abundance level.
- RTS-MS3 close-out modes (5 peptides per trigger) minimized background in blank channels, improving accuracy for low-abundance knockout signals.
Comparative Depth
- SPS-MS3 detected approximately 400 fewer protein groups than RTS-MS3 under identical acquisition times.
- Combination of RTS with COMET search in-line enhanced the number of quantified proteins by an additional 7% over post-acquisition analysis.
Benefits and Practical Applications
This TMTpro 18plex standard serves as a versatile benchmark for:
- Calibrating instrument methods to achieve optimal quantitation performance.
- Comparing acquisition strategies across different platforms and laboratories.
- Routine quality control to detect drift in mass accuracy, resolution, or quantitation precision.
- Method development for complex sample multiplexing in proteomics research, clinical biomarker studies, and biopharma QC.
Future Trends and Opportunities
Expanding on this work, next-generation standards could:
- Integrate targeted peptides from diverse organisms or post-translational modifications to assess site-specific quantitation.
- Leverage advanced real-time decision engines incorporating machine learning to optimize acquisition on the fly.
- Adapt multiplex standards for emerging platforms such as trapped ion mobility spectrometry (TIMS) or parallel reaction monitoring (PRM).
- Facilitate multi-laboratory ring trials to set community guidelines for multiplexed proteomics QA/QC.
Conclusion
The prototype TMTpro 18plex yeast digest standard effectively benchmarks quantitative accuracy and precision across MS2, SPS-MS3, and RTS-MS3 workflows. RTS-MS3 methods consistently yield higher proteome coverage and improved measurement fidelity, particularly for low-abundance channels. Adoption of this standard enables rigorous method validation, supports reproducible multi-site studies, and accelerates innovation in multiplexed proteomics.
References
- Paulo J.A., O’Connell J.D., Gygi S.P. A triple knockout (TKO) proteomics standard for diagnosing ion interference in isobaric labeling experiments. J. Am. Soc. Mass Spectrom. 2016;27(10):1620–1625.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
A TMTpro 18plex Proteomics Standard for Assessing Protein Measurement Accuracy and Precision 
2021|Thermo Fisher Scientific|Posters
A TMTpro 18plex Proteomics Standard for Assessing Protein Measurement Accuracy and Precision Jae Choi1, Yan Liu2, Rosa Viner2, Joao A. Paulo3, Ryan Bomgarden1 1Thermo Fisher Scientific, Rockford, IL; 2Thermo Fisher Scientific, San Jose, CA; 3Harvard Medical School, Boston, MA ABSTRACT…
Key words
rts, rtsprotein, proteinidentifications, identificationsgroups, groupstribrid, tribridagc, agcparental, parentaleclipse, eclipsepsms, psmsorbitrap, orbitrapstrains, strainsscan, scanquan, quanicsw, icswcid
Implementing Comet search engine into Proteome Discoverer to improve TMT Real-Time Search data processing
2021|Thermo Fisher Scientific|Posters
Implementing Comet search engine into Proteome Discoverer to improve TMT Real-Time Search data processing 1 Yang Liu, 2Frank Berg, 1 William D. Barshop, 1Jesse D. Canterbury, 1David Horn, 1David Bergen, 1Romain Huguet, 1Rosa Viner 1Thermo Fisher Scientific, San Jose, CA,…
Key words
comet, cometpercolator, percolatorsequest, sequestdecoy, decoysearch, searchpsm, psmdiscoverer, discovererproteome, proteomerts, rtsfixvalue, fixvalueprotein, proteinfix, fixquan, quansequestht, sequesthttribrid
Real-Time Search enables a new gold standard for TMT quantitation accuracy on the Orbitrap Eclipse Tribrid mass spectrometer
2020|Thermo Fisher Scientific|Applications
APPLICATION NOTE 65729 Real-Time Search enables a new gold standard for TMT quantitation accuracy on the Orbitrap Eclipse Tribrid mass spectrometer Authors: Aaron M. Robitaille1, Romain Huguet1, Ryan Bomgarden2, Jesse D. Canterbury1, Daniel Lopez-Ferrer1 Thermo Fisher Scientific, San Jose, CA…
Key words
search, searchtmt, tmtreal, realcomet, cometxcorr, xcorrsps, spstribrid, tribridorbitrap, orbitrapmass, masstime, timeeclipse, eclipsequantitation, quantitationinterference, interferenceengine, engineaccuracy
A Modified Orbitrap Tribrid Mass Spectrometer with Real Time Search and Advanced Spectral Processing Enhances Multiplexed Proteome Coverage and Quantification Accuracy
2019|Thermo Fisher Scientific|Posters
POSTER NOTE 65536 A Modified Orbitrap Tribrid A Modified Orbitrap Tribrid Mass Spectrometer with Real Time Search and Advanced Spectral Processing EnhancesCov Multiplexed Proteome Multiplexed ProteomeAaron Coverage and1,Quantification M. Robitaille Romain Huguet 1, Accuracy Derek J. Bailey 1, G 1…
Key words
sps, spstmt, tmttribrid, tribridrts, rtssearch, searchreal, realmodified, modifiedauto, autoorbitrap, orbitrapspssps, spsspsinterference, interferencespectrometer, spectrometertime, timeion, ionmass