Best Practices for Addressing Problems Associated With Unstable Solvents in an (U)HPLC Environment
Technical notes | 2018 | Agilent TechnologiesInstrumentation
Solvent stability is critical in (U)HPLC operations to prevent formation of corrosive byproducts that can damage separation columns and instrument components. Reactive species such as halides, peroxides and strong acids may form under non-ideal storage and handling conditions, posing safety and maintenance challenges.
This technote aims to outline best practices for handling unstable solvents in a (U)HPLC setting, focusing on operational measures to avoid instrument deterioration, maintain analytical performance and ensure laboratory safety.
The guidance is based on evaluation of solvent degradation pathways and their impact on hardware. Recommendations include inert-gas blanketing, use of stabilized reagents, appropriate container sizes and dedicated solvent lines.
Proactive control of moisture, light and air in solvent reservoirs minimizes formation of hydroperoxides and free radicals. Small solvent bottles (≤1 L) and low-flow purging during idle times reduce solvent aging. Reactive solvents require bypassing or periodic flushing of solvent selection valves and degasser chambers. Material compatibility is crucial: PTFE seals may be replaced with PE, sapphire pistons can substitute ceramic ones in acidic organic mixtures, and DMSO exposure demands timely leak-sensor replacement.
Implementing these practices extends instrument longevity, reduces unexpected maintenance, and enhances data reproducibility. Laboratories can safely employ a wider range of solvent systems, including halogenated solvents, ethers and buffered mobile phases, by selecting proper stabilizers and materials.
The shift toward greener and safer solvents is accelerating, with emerging alternatives like 2-MeTHF, MTBE and cyclopentyl methyl ether replacing traditional ethers. Vendors are expanding portfolios of sustainable solvents. Continued innovation in inert materials and inline stabilization technologies will further improve operational safety and environmental compliance.
Effectively managing solvent stability in (U)HPLC processes requires a combination of preventive storage conditions, compatible hardware materials and proper solvent selection. Adopting these best practices ensures instrument integrity, analytical consistency and a safer laboratory environment.
HPLC
IndustriesManufacturerAgilent Technologies
Summary
Importance of the Topic
Solvent stability is critical in (U)HPLC operations to prevent formation of corrosive byproducts that can damage separation columns and instrument components. Reactive species such as halides, peroxides and strong acids may form under non-ideal storage and handling conditions, posing safety and maintenance challenges.
Study Objectives and Overview
This technote aims to outline best practices for handling unstable solvents in a (U)HPLC setting, focusing on operational measures to avoid instrument deterioration, maintain analytical performance and ensure laboratory safety.
Methodology and Used Instrumentation
The guidance is based on evaluation of solvent degradation pathways and their impact on hardware. Recommendations include inert-gas blanketing, use of stabilized reagents, appropriate container sizes and dedicated solvent lines.
- UHPLC systems with stainless steel flow paths (Agilent 1290 Infinity series, 1260 Infinity II pumps).
- Degassing units for removing gas bubbles and preventing condensation of volatile components.
- Sustainable inlet and outlet valves with PEEK or polyimide seals.
- Samplers with robust needle seats and leak sensors.
Main Findings and Discussion
Proactive control of moisture, light and air in solvent reservoirs minimizes formation of hydroperoxides and free radicals. Small solvent bottles (≤1 L) and low-flow purging during idle times reduce solvent aging. Reactive solvents require bypassing or periodic flushing of solvent selection valves and degasser chambers. Material compatibility is crucial: PTFE seals may be replaced with PE, sapphire pistons can substitute ceramic ones in acidic organic mixtures, and DMSO exposure demands timely leak-sensor replacement.
Benefits and Practical Applications
Implementing these practices extends instrument longevity, reduces unexpected maintenance, and enhances data reproducibility. Laboratories can safely employ a wider range of solvent systems, including halogenated solvents, ethers and buffered mobile phases, by selecting proper stabilizers and materials.
Future Trends and Potential Uses
The shift toward greener and safer solvents is accelerating, with emerging alternatives like 2-MeTHF, MTBE and cyclopentyl methyl ether replacing traditional ethers. Vendors are expanding portfolios of sustainable solvents. Continued innovation in inert materials and inline stabilization technologies will further improve operational safety and environmental compliance.
Conclusion
Effectively managing solvent stability in (U)HPLC processes requires a combination of preventive storage conditions, compatible hardware materials and proper solvent selection. Adopting these best practices ensures instrument integrity, analytical consistency and a safer laboratory environment.
References
- Sigma Aldrich. Stabilizer Systems for Solvents. Technical Note.
- Agilent Technologies. Best Practices for Addressing Problems Associated With Unstable Solvents in (U)HPLC. TechNote 01200-90092.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Thermo Scientific Dionex UltiMate 3000 Series Nano/Cap System NCS-3500RS Nano/Cap Pump NCP-3200RS Operating Instructions
2016|Thermo Fisher Scientific|Manuals
Thermo Scientific Dionex UltiMate 3000 Series Nano/Cap System NCS-3500RS Nano/Cap Pump NCP-3200RS Operating Instructions (Original Operating Instructions) Revision: 2.0 Date: January 2016 © 2016 Thermo Fisher Scientific Inc. Doc No. 4820.4101 UltiMate 3000 Series: NCS-3500RS and NCP-3200RS Operating Instructions UltiMate…
Key words
instructions, instructionspump, pumpoperating, operatingmeter, metermodule, moduleflow, flowseal, sealhead, headpressure, pressurevalve, valveloading, loadingproflow, proflowchromeleon, chromeleonpurge, purgeremedial
Best Practices for Using an Agilent LC System - Technical Note
2022|Agilent Technologies|Technical notes
Best Practices for Using an Agilent LC System Technical Note This technical note describes best practices for using an Agilent LC system. Housekeeping 2 Daily / Weekly Tasks 5 Power Up / Shut Down the System Recommendations for Degassers Recommendations…
Key words
recommendations, recommendationspractices, practicesbest, bestnote, notetechnical, technicalpumps, pumpsseal, sealsamplers, samplerswash, washhousekeeping, housekeepingdamage, damagedegassers, degassersinline, inlinesolvent, solventmultiwash
Around the LC System in 45 Slides
2016|Agilent Technologies|Presentations
Around the LC System in 45 Slides Rita Steed and Alex Ucci Pressure, Peak Shape, & RT Problems Plan Ahead for a Smooth Trip Solvent Bottle • What can you do to reduce or anticipate potential problems with your chromatography?…
Key words
confidentiality, confidentialityseal, sealvalve, valvelabel, labelphase, phasepump, pumpcare, carecolumn, columnless, lesscaptiva, captivamobile, mobilepressure, pressuresample, samplesolvent, solventwear
Goog HPLC practice guide
2021|KNAUER|Guides
Science with Passion HPLC Support: International Support: Contact your local KNAUER partner for support: www.knauer.net/en/Support/Distributors-worldwide Support in Germany (Austria & Switzerland on case-to-case basis): Phone: +49 30 809727-111 (workdays 9-17h CET) Fax : +49 30 8015010 Email: [email protected] Publisher: KNAUER…
Key words
practice, practiceguide, guidehplc, hplcgood, goodether, ethergloves, glovesmobile, mobilesolvents, solventschromophore, chromophorecutoff, cutoffhow, howpropanol, propanolphase, phasetrichloroethylen, trichloroethylenethyl