Manufacturer
KNAUER
KNAUER
Based in Berlin, KNAUER is a medium-sized, owner-managed company that has been serving the sciences since 1962. We develop and manufacture scientific instruments of superior quality for liquid chromatography, including: Analytical HPLC/UHPLC, Preparative HPLC, Fast protein liquid chromatography (FPLC), Multi-column chromatography/Simulated moving bed (SMB) chromatography. High pressure dosing, in-line detection systems, and osmometry are additional business areas.
Distributor
KNAUER
KNAUER manufactures high-end scientific instruments for research, routine analysis, quality assurance, and other applications.
KNAUER High Precision AZURA P 6.1L HPLC PumpKNAUER AZURA P 6.1L HPLC pump with up to 1000 bar, precise gradient control, and low pulsation for reliable UHPLC workflows.
Features

KNAUER High Precision AZURA P 6.1L HPLC Pump

The KNAUER AZURA P 6.1L is a high-precision analytical HPLC pump designed for accurate and reproducible solvent delivery in both HPLC and UHPLC applications.

The pump combines essential features of a modern pumping system, including low pulsation, high pressure capability, precise gradient formation, and flexible configuration. It is available in both HPG (high-pressure gradient) and LPG (low-pressure gradient) versions, with a flow range up to 50 mL/min.

The system uses a dual serial piston mechanism with active pulsation compensation, ensuring excellent flow stability under all conditions. A low-dead-volume microfluidic mixer enables efficient and precise solvent mixing.

Different pump heads are available:

  • up to 10 mL/min at 862 bar
  • up to 5 mL/min at 1000 bar

Optional configurations include biocompatible materials (PEEK, ceramic, titanium) and specialized setups for normal phase applications (e.g., hexane, heptane).

Key Features

  • High-precision analytical HPLC/UHPLC pump
  • Pressure up to 1000 bar
  • Flow rates up to 50 mL/min
  • Dual piston system with low pulsation
  • Active pulsation compensation
  • HPG and LPG gradient configurations
  • Microfluidic mixer (50–600 µl)
  • RFID pump head detection
  • Optional degasser
  • Biocompatible configurations available
  • Robust operation with demanding solvents

Technical Specifications

Solvent Delivery

  • Pump type: analytical HPLC pump
  • Mechanism: dual serial piston
  • Flow range: 0.001 – 50 mL/min
  • Max pressure: up to 1000 bar
  • Flow accuracy: ± 0.25%
  • Precision: ≤ 0.04% RSD

Gradient

  • Range: 0–100% (0.1% steps)
  • HPG accuracy: ± 0.3%
  • HPG precision: < 0.1% RSD
  • LPG accuracy: ± 0.3% to ± 2%
  • LPG precision: < 0.5% RSD

Mixing

  • Microfluidic mixer
  • Volumes: 50 / 100 / 200 / 400 / 600 µl
  • Low dead volume

Degasser (optional)

  • 2–4 channels
  • Max flow: 10 mL/min per channel
  • Residual O₂: < 0.5 ppm
  • Membrane: fluoropolymer

Materials

  • PEEK, Tefzel®, Systec AF
  • Optional biocompatible materials (titanium, ceramic)

Control & Communication

  • LAN, analog inputs
  • Event outputs (TTL, relay)
  • Optional mobile control
  • RFID pump head detection

Operating Conditions

  • Temperature: 4–40 °C
  • Humidity: < 90% non-condensing

Dimensions & Power

  • Dimensions: 361 × 208 × 523 mm
  • Weight: 14.1 kg
  • Power supply: 100–240 V, 50–60 Hz
  • Power consumption: max 100 W
Manufacturer
KNAUER
KNAUER
Based in Berlin, KNAUER is a medium-sized, owner-managed company that has been serving the sciences since 1962. We develop and manufacture scientific instruments of superior quality for liquid chromatography, including: Analytical HPLC/UHPLC, Preparative HPLC, Fast protein liquid chromatography (FPLC), Multi-column chromatography/Simulated moving bed (SMB) chromatography. High pressure dosing, in-line detection systems, and osmometry are additional business areas.
Distributor
KNAUER
KNAUER manufactures high-end scientific instruments for research, routine analysis, quality assurance, and other applications.
 

News

Toggle
Oligos Made Easy - Part 9: 2D-LC
Article | Academy

Oligos Made Easy - Part 9: 2D-LC

Discover how 2D-LC enhances oligonucleotide analysis by improving resolution, separating complex impurities, and enabling deeper insight beyond conventional 1D-LC methods.
KNAUER
tag
share
more
News from LabRulezLCMS Library - Week 10, 2026
Article | Application

News from LabRulezLCMS Library - Week 10, 2026

This week we bring you application notes by KNAUER, Shimadzu, Thermo Fisher Scientific and Waters Corporation and technical note by Agilent Technologies!
LabRulez
tag
share
more
Other projects
GCMS
ICPMS
Follow us
FacebookX (Twitter)LinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike